Before contacting for repairs
8
. Before contacting for repairs
Checking causes and countermeasures
The following problems can occur even if the operating table is not malfunctioning. Check the following points
before requesting repairs.
Status
Possible cause
Measures
The table cannot be
turned on.
The control unit connector is not
connected properly.
Insert the connector completely.
(
à
Page 15)
The battery is fully discharged.
Charge the battery.
(
à
Page 16)
A switch on the control
unit does not function.
The control unit connector is not
connected properly.
Insert the connector completely.
(
à
Page 15)
E switch is not pressed simultaneously
with the function switch.
Hold down E switch and press the switch.
(
à
Page 19 to 25)
The brake switch on
the control unit does
not function.
The emergency brake release handle is
in the release (UNLOCK) position.
Turn the emergency brake release
handle toward "LOCK". (
à
Page 36)
If the situation does not improve even if the above countermeasures are implemented, request repairs from
your distributor or Mizuho.