![Mizuho MOT-1602BW Operator'S Manual Download Page 38](http://html1.mh-extra.com/html/mizuho/mot-1602bw/mot-1602bw_operators-manual_1812822038.webp)
36
7
T
roubleshooting
7
. Troubleshooting
When the control unit cannot be used
z
Use the emergency brake release handle to release the
brake
WARNING
Do not tilt the tabletop while the emergency brake
release handle is in the "UNLOCK" position.
The patient may fall from the operating table.
In case of electrical trouble, the operating table can be moved by using the
emergency brake release handle.
Follow the procedure below to release the brake.
1.
Turn the emergency brake release handle counterclockwise
(to the left).
The brake is released.
Turn the handle
CAUTION
To move the operating table after you released
the brake, activate the brake again with the
following procedure. If the operating table is
moved with the brake released, the patient may
fall from the operating table.
1. Return the emergency brake release handle
clockwise (to the right).
Turn the handle
2. Hold
ENERGIZE
and press
on the control unit.
The operating table is fixed.