Electromagnetic Compatibility
App.
Electromagnetic Compatibility
Medical Electrical Equipment needs special precautions regarding EMC and needs to be installed and put
into service according to the EMC information provided in this manual.
Portable and mobile RF communications equipment can affect Medical Electrical Equipment.
The use of Accessories, transducers, and cables other than those specified, with the exception of transducers
and cables sold by the Manufacturer of this device as replacement parts for internal components, may result
in increased Emissions or decreased Immunity of Operating Table MOT-1602BW.
Operating Table MOT-1602BW should not be used adjacent to or stacked with other equipment and that if
adjacent or stacked use is necessary, Operating Table MOT-1602BW should be observed to verify normal
operation in the configuration in which it will be used.
GUIDANCE AND MANUFACTURER’S DECLARATION – ELECTROMAGNETIC EMISSIONS
Operating Table MOT-1602BW is intended for use in the electromagnetic environment specified below.
The customer or the user of Operating Table MOT-1602BW should assure that it is used in such an
environment.
Emissions test
Compliance
Electromagnetic environment – guidance
RF emissions CISPR 11
Group 1
Operating Table MOT-1602BW uses RF energy only for its
internal function. Therefore, its RF emissions are very low and
are not likely to cause any interference in nearby electronic
equipment.
RF emissions CISPR 11
Class A
Operating Table MOT-1602BW is suitable for use in all
establishments, other than domestic establishments and those
directly connected to the public low-voltage power supply
network that supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations /
flicker emissions
IEC 61000-3-3
Complies