5
2
Safety precaution
Other
WARNING
●
Prohibited
•
Do not disassemble and/or modify the operating table. Otherwise, malfunction may
occur.
●
Other medical electrical equipment to be used together with the operating table
•
Before use, check that the operating table does not malfunction due to
electromagnetic interference from the equipment.
Medical electrical equipment to be used together with the operating table may
generate electromagnetic interference, which may result in malfunctioning of the
operating table.
• When using a high-frequency surgical equipment and/or a cardiac defibrillator, refer
to their operator's manual provided by the manufacturers. Improper usage may
cause the operator and the patient to get burned and/or devices to malfunction.
●
Patient position
•
When using the tabletop or accessories to secure a patient’s body position, always
observe the patient’s condition.
•
Neuroparalysis may occur to the patient.
●
Allowable load
•
Do not apply a load which exceeds the allowable load*. The operating table may not
function, which may result in failures.
* Lift: 360 kg / Lifts other than: 270 kg
●
Preventive maintenance and inspections
•
Make sure to inspect and maintain the operating table before and after use. The
operating table may require replacement of the parts due to significant wear,
deterioration, and/or breakage depending on the usage conditions and frequency
usage.
•
For preventive maintenance and inspections, contact your distributor or Mizuho
directly.
●
The usage of high-frequency surgical equipment
• When using a high-frequency surgical equipment and/or a cardiac defibrillator,
refer to the operator's manual provided by the manufacturers. Improper use causes
burn injuries to he operator and the patient.
●
Antistatic measure
• Do not use the operating table on floors and/or with accessories that do not
possess static electricity countermeasures. This may impede surgical operations.