6 Maintenance and Troubleshooting
No. 99MAM029A
6.2
Error Displays
6.2.1 Error Displays with Power LED
When the power is turned on, the power LED is lit on when the product is in the normal
state. If it is blinking, there may the errors shown below. Address it according to the error
contents.
Power LED blinking pattern State of the Product
Data input Handling method
Repetition of blinking at 0.6 s
on and 0.6 s off.
Abnormally high volt-
age
Not possible Turn off the power
and turn it on again.
Repetition of blinking at 1.5 s
off, 0.3 s on, 0.3 s off and 0.3
on
Voltage drop alarm
(when the voltage
drops and the remain-
ing battery level be-
comes low)
Possible
It is recovered
when the voltage
returns to the nor-
mal range.
(Replace batteries)
Repetition of blinking at 0.6 s
on and 0.6 s off.
Abnormally low voltage
(when voltage drops
and operation be-
comes impossible)
Not possible Turn off the power
and turn it on again.
Tips
• When the power of the product is turned off, all the measurement data automatically
saved for the statistical calculation up to that point of time is deleted. Also, if the AC
adapter is connected/disconnected while the product is operating, all the measurement
data automatically saved for the statistical calculation up to that point of time is lost, re-
gardless of whether the product is powered with the AC adapter or not. Please handle
with care.
• When operating the product with batteries, if the room temperature drops below 10 °C,
the battery life becomes considerably shorter. Use an AC adapter when the room tem-
perature is below 10 °C.
• When the batteries are consumed, the power LED may malfunction as the voltage
becomes below the operating voltage. If the power LED starts blinking, change the
batteries as promptly as possible or switch to the AC adapter.