![Mitutoyo DP-1VA LOGGER User Manual Download Page 46](http://html1.mh-extra.com/html/mitutoyo/dp-1va-logger/dp-1va-logger_user-manual_1812411046.webp)
30
3 Advanced Operations and Useful Functions
No. 99MAM029A
5
Press
DATA
repeatedly until the param-
eter settings list is printed.
» Parameter setup mode is completed.
2
Set the tolerance judgment.
For further details on setting and selection method of the tolerance judgment, refer to
"2.3.1
Tolerance Settings" on page 19.
3
Input the measurement data.
1
Press
DATA
.
» The measurement data is printed in
D chart format each time
DATA
is
pressed.
D chart
Symbol for the
outside toler-
ance
Tips
• Symbol
▼
▼
in the D chart represents that the measurement data is out of the toler-
ance.
• The maximum number of measurement data that can be handled with MODE2 is
9,999.
• When pressing
STAT.
OUT LOG
after inputting the measurement data, the statistical calculation
value and histogram are printed as in the case of MODE1. Once 9,999 measurement
data are input, the statistical calculation result is automatically printed.
POWER
TOL.
REC/STOP
PRINTER
CLEAR
STAT.
OUT LOG
CANCEL
DATA
FEED
Summary of Contents for DP-1VA LOGGER
Page 16: ...xiv No 99MAM029A...
Page 22: ...6 1 Outline No 99MAM029A...
Page 42: ...26 2 Basic Operations No 99MAM029A...
Page 76: ...60 4 Output No 99MAM029A...
Page 88: ...72 6 Maintenance and Troubleshooting No 99MAM029A...