6. Packing Accessory 

 

1. 

Draining pipe 

1 pc 

2. 

Mesh tray   

2-3 pcs 

3. 

Plug 

1 pc 

4. 

Bracket     

1 pc 

5. 

Manual holder 

1 pc 

6. 

Fuse   

2 pcs 

7. 

Operation manual   

1 pc 

8. 

Measuring cup 

1 pc 

9. 

Sealing ring 

1 pc 

 

7. Display and keypad description 

7.1 Control Panel 

1) Pressure display 

Display on time the air pressure inside the 
chamber. Unit:KPa

 

2) Temperature display 

Display  on  time  the  temperature  inside the  cavity. 

Unit: ºC

 

3) Status/ Error code number display 

Applied  for  displaying  kinds  of  status  of  the 
autoclave during its working time. See more details 

in “Appendix working status display form”.

 

When  alarm  occurs  to  the  autoclave,  the  digital 
pipe  of  the  display  will  show  the  alarm  code 
number,  the  user  can  find  out  the  malfunction 
part.(Once  the  malfunction  occurring  to  the 
autoclave, please contact the service 
center or distributor immediately).

 

4) UNWRAPPED button 

It

s used to the sterilization 

choice especially for the unwrapped 
devices. The working temperature 
is 121

º

C or 134

º

C. 

After having been selected, its indicator   
light is on. 

 

5) WRAPPED    button 

It

s used to be the sterilization choice especially for 

Summary of Contents for SUN12-II

Page 1: ...on Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN8 SUN12 SUN16 SUN18 SUN23 Steam Sterilizer File No MTQM C1 7 5 1 0...

Page 2: ...am Specification 2 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 5 8 Emergency Condition 8 9 Maintenance and Service 9 10 Transportation and Storage 11 11 Attention 12 Append...

Page 3: ...n Scope This machine can be applied in sterilizing for the tooth device operation device Sanitary Material and other heat resisting articles existing in the Dept of Stomatology Ophthalmology and lab T...

Page 4: ...temperature of operate 0 40 C Type of sterilization cycles Class B Wire heat resistance level 70 C N W 23L 53kg N W 18L 48kg N W 16L 47kg N W 12L 46kg N W 8L 44kg 4 Diagram Specification Symbol Meanin...

Page 5: ...onment temperature 5 40 Relative humidity 85 barometric pressure 70kPa 106kPa minimum brightness 215 15 lx 5 3 The Steam Sterilizer shall be installed on the horizontal working base Put machine on the...

Page 6: ...ay contact you local supply authority for further information Hinweis Beim Anlaufen Starten dieser Ausr stungen kann ein kurzzeitiger Spannungseinbruch auftreten insbesondere bei schlechter Netzqualit...

Page 7: ...for displaying kinds of status of the autoclave during its working time See more details in Appendix working status display form When alarm occurs to the autoclave the digital pipe of the display will...

Page 8: ...ndicator for alarming the shortage of water in the distilled water tank and reminding for filling distilled water 11 Light indicator waste water tank full It s the light indicator for alarming the ful...

Page 9: ...umping open the drain valve to drain the exhaust 0 0kPa to close drain valve intermittent steam generator to pump water to generate steam when the pot the temperature reaches the set value the end of...

Page 10: ...e provide accessories to take devices Must use handheld to take things prevent burns 8 Emergency Condition During the working time the Steam Sterilizer will automatically alarming then release pressur...

Page 11: ...to the right MAX position 7 E7 long du Fault of temperature senor on steam generator Inspecting the temperature sensor on steam generator 8 E8 long du Fail on pressure rising Inspect water flowing ro...

Page 12: ...ommendation by the supplier 9 5 The replacement work for sealing ring Tool One flat screwdriver No sharp edge at the head 1 One hand catches the lip of the sealing ring the other hand insert the screw...

Page 13: ...al components must cut off power supply first And only the manufacturer or professional person designated by the manufacturer can operate 10 Transportation and Storage 10 1 The preparatory work before...

Page 14: ...e door of the Steam Sterilizer in case of scald when open the door 9 Shut off the power before discharging installing the sealing ring and the work shall be carried out after further sufficient coolin...

Page 15: ...13 Appendix 1 Steam loop Work Description...

Page 16: ...14 2 Circuit diagram of sterilizer...

Page 17: ...playing Ed B D test when put the test material onto the tray inside the chamber firstly turn on the power by the green switch close the door keeping press the UNWRAPPED button around 10s When the boar...

Reviews: