Vacuum test and "B&D" test methods
Vacuum test
: connecting well the pressure meter(if necessary) from the backing
test connector, turn on the power by the green switch, close the door, keeping
press the temperature choice button around 10s, the board would displaying "nn",
and then it would start the vacuum testing automatically. After vacuum pulse, it
would holding the vacuum status and end by displaying "Ed".
B&D test:
when put the test material onto the tray inside the chamber firstly, turn
on the power by the green switch, close the door, keeping press the
"UNWRAPPED" button around 10s, When the board displaying "bd", then press
start/stop button, the autoclave would start the BD test automatically and end by
displaying "Ed".