Installation
5.1 Remove the packing
Dismantle thin film, open the packing,get the equipment from the packing
Hauling precautions:
·
forbid lift up door
·
forbid lift up legs of machine
·
forbid Rollover or upside down
·
Hauling sketch map is as follows:
5.2 The air-cooling window attached to the exterior surface of the Steam Sterilizer
shall be non-clog by dust or articles, and the Steam Sterilizer shall working in the
proper air flowing environment,and n
o dangerous flammable and explosive gas.
Equipment normal working conditions
:
environment temperature
:
+5
~
+40℃
Relative humidity
:
≤85
%
barometric pressure
:
70kPa
~
106kPa
minimum brightness: 215
±
15 lx,
5.3 The Steam Sterilizer shall be installed on the horizontal working base. Put
machine on the table, refer to specifications, table must be able to bear the weight
of the machine.
The clearance between the outside of Steam Sterilizer and the
around articles shall be 10cm as retained, and the top surface shall be 20cm
retained.
Used to exhaust or drain.
5.4 The 2 front feet of Steam Sterilizer can be adjusted the height, please ensure
the front of machine slightly higher than the back.
5.5 The machine will be hot during working time, keep it away from any
combustible materials!
The equipment must be far away from flammable items