MingTai SUN12-II Operation Manual Download Page 10

 

 

When the drying time is reduced to zero, turn off the exhaust, dry intake valve 
opens, so that the pressure to achieve and external balance, open the automatic 
doors, complete sterilization.   

7.3.6 Sterilization program is as follows: 

 

 

 

134

 

UNWRAPPED 

121

 

UNWRAPPED 

134

 

WRAPPED 

121

 

WRAPPED 

121

 

COTTON 

121

 

PLASTIC 

Sterilizing 
temperature 

134

 

121

 

134

 

121

 

121

 

121

 

Sterilizing 
pressure 

210 

110 

210 

110 

110 

110 

Vacuum times 

Sterilizing  
time 

20 

20 

20 

20 

Drying time 

15 

20 

15 

20 

20 

20 

 
7.4 Emergency exit, Long press the start key 5 s, Machine will out of sterilization procedures, 

meanwhile 

"EC" alarm 

7.5 After sterilization, use handheld we provide accessories    to take devices 

  Must use handheld to take things, prevent burns!

 

8. Emergency Condition 

During  the  working  time,  the  Steam  Sterilizer  will automatically  alarming, then 
release  pressure,  stop  heating  and  display  the  error  code  (The  definition  of  the 
error  code  is  as  the  below  table:  error  code  and  resolving  method  ).  It  would 
ensure the safety of the operator if the emergency occur. 
Any  emergency  happen,  please  check  the  warning  code  table  for  resolving  the 
according malfunctions. If not possible to resolve, please contact the dealers.  We 
will try our best to provide help to you in short time. 
  After  the  alarm, The machine  can't  work,  Long  press  the  start  key,  can  unlock 
alarm. 

 

Must 

wait  for  the  pressure  display  “0

  Kpa

”  can  open  the  door

 

  there will be set aside heat and residual hot water, Operator also need keep 

away from it , prevent burns! 
 
 
 

Summary of Contents for SUN12-II

Page 1: ...on Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN8 SUN12 SUN16 SUN18 SUN23 Steam Sterilizer File No MTQM C1 7 5 1 0...

Page 2: ...am Specification 2 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 5 8 Emergency Condition 8 9 Maintenance and Service 9 10 Transportation and Storage 11 11 Attention 12 Append...

Page 3: ...n Scope This machine can be applied in sterilizing for the tooth device operation device Sanitary Material and other heat resisting articles existing in the Dept of Stomatology Ophthalmology and lab T...

Page 4: ...temperature of operate 0 40 C Type of sterilization cycles Class B Wire heat resistance level 70 C N W 23L 53kg N W 18L 48kg N W 16L 47kg N W 12L 46kg N W 8L 44kg 4 Diagram Specification Symbol Meanin...

Page 5: ...onment temperature 5 40 Relative humidity 85 barometric pressure 70kPa 106kPa minimum brightness 215 15 lx 5 3 The Steam Sterilizer shall be installed on the horizontal working base Put machine on the...

Page 6: ...ay contact you local supply authority for further information Hinweis Beim Anlaufen Starten dieser Ausr stungen kann ein kurzzeitiger Spannungseinbruch auftreten insbesondere bei schlechter Netzqualit...

Page 7: ...for displaying kinds of status of the autoclave during its working time See more details in Appendix working status display form When alarm occurs to the autoclave the digital pipe of the display will...

Page 8: ...ndicator for alarming the shortage of water in the distilled water tank and reminding for filling distilled water 11 Light indicator waste water tank full It s the light indicator for alarming the ful...

Page 9: ...umping open the drain valve to drain the exhaust 0 0kPa to close drain valve intermittent steam generator to pump water to generate steam when the pot the temperature reaches the set value the end of...

Page 10: ...e provide accessories to take devices Must use handheld to take things prevent burns 8 Emergency Condition During the working time the Steam Sterilizer will automatically alarming then release pressur...

Page 11: ...to the right MAX position 7 E7 long du Fault of temperature senor on steam generator Inspecting the temperature sensor on steam generator 8 E8 long du Fail on pressure rising Inspect water flowing ro...

Page 12: ...ommendation by the supplier 9 5 The replacement work for sealing ring Tool One flat screwdriver No sharp edge at the head 1 One hand catches the lip of the sealing ring the other hand insert the screw...

Page 13: ...al components must cut off power supply first And only the manufacturer or professional person designated by the manufacturer can operate 10 Transportation and Storage 10 1 The preparatory work before...

Page 14: ...e door of the Steam Sterilizer in case of scald when open the door 9 Shut off the power before discharging installing the sealing ring and the work shall be carried out after further sufficient coolin...

Page 15: ...13 Appendix 1 Steam loop Work Description...

Page 16: ...14 2 Circuit diagram of sterilizer...

Page 17: ...playing Ed B D test when put the test material onto the tray inside the chamber firstly turn on the power by the green switch close the door keeping press the UNWRAPPED button around 10s When the boar...

Reviews: