MingTai SUN12-II Operation Manual Download Page 13

 

 

11 

2)When  the  pressure  in  the  sterilizer  reaches  100  kpa

Pull  rings  on  safety 

valve,  make  the  safety  valve  open,about  2S,    if  steam  jet,  the  safety  valv
e  work  is  normal.  Ifnot,  the  safety  valve  need  for  effective  inspection  or 
replacement

 

  Pull  rings  on  safety  valve  will  steam  jet,  It  is  best  to  use  tools  such 

as  screwdriver  when  pull.  Do  not  use  finger  directly,  Operator  also  need  ke
ep  away  from  it  ,  prevent  burns! 
 
Replace  the  safety  valve: 

This  operation  is  limited  to  the  professionals 

1)  Loosen  the  set  screws  of  the  safety  valve,  take  the  safety  valve  down  f
rom  the  panel, 
2)  replace  with  a  qualified  safety  valve  (must  provide  by  manufacturers). 
3)  Test  a  sterilization  process 

The machine maintenance and replacement of electrical components must cut 

off  power  supply  first.  And  only  the  manufacturer  or  professional  person 
designated by the manufacturer can operate 
 

10. Transportation and Storage

 

10.1 The preparatory work before 
transportation and storage: 
Shut off the power, draw out the plug and cool the Steam Sterilizer down. 
 
10.2 Draining   
Empty  the  water  storing  tank  after  the  machine  become  colling;  use  the  pipe 
packing together with the machine, connecting to the valve showing in the above 
picture 4 or 3, 4 is the water exiting mouth for used water collecting tank, 3 is the 
water  exiting  mouth  of  the  distilled  source  water  storing  tank.  Revolving  the 
water exit button by anticlockwise direction so as to release the water. 
 
10.3  The  Steam  Sterilizer  shall  be  transferred  and  stored  with  the  below 
conditions: 
Ambiance temperature: -

40ºC 

-

+55ºC

 

Relative humidity: 

  85% 

Atmospheric pressure: 500HPA-1060HPA 
 

   

Note: Do not drag or drop during transportation.

   

 

 

Summary of Contents for SUN12-II

Page 1: ...on Manual Please read the manual carefully operate it according to the manual strictly and avoid any hazard This manual applies to SUN8 SUN12 SUN16 SUN18 SUN23 Steam Sterilizer File No MTQM C1 7 5 1 0...

Page 2: ...am Specification 2 5 Installation 3 6 Packing accessory 5 7 Display and keypad description 5 8 Emergency Condition 8 9 Maintenance and Service 9 10 Transportation and Storage 11 11 Attention 12 Append...

Page 3: ...n Scope This machine can be applied in sterilizing for the tooth device operation device Sanitary Material and other heat resisting articles existing in the Dept of Stomatology Ophthalmology and lab T...

Page 4: ...temperature of operate 0 40 C Type of sterilization cycles Class B Wire heat resistance level 70 C N W 23L 53kg N W 18L 48kg N W 16L 47kg N W 12L 46kg N W 8L 44kg 4 Diagram Specification Symbol Meanin...

Page 5: ...onment temperature 5 40 Relative humidity 85 barometric pressure 70kPa 106kPa minimum brightness 215 15 lx 5 3 The Steam Sterilizer shall be installed on the horizontal working base Put machine on the...

Page 6: ...ay contact you local supply authority for further information Hinweis Beim Anlaufen Starten dieser Ausr stungen kann ein kurzzeitiger Spannungseinbruch auftreten insbesondere bei schlechter Netzqualit...

Page 7: ...for displaying kinds of status of the autoclave during its working time See more details in Appendix working status display form When alarm occurs to the autoclave the digital pipe of the display will...

Page 8: ...ndicator for alarming the shortage of water in the distilled water tank and reminding for filling distilled water 11 Light indicator waste water tank full It s the light indicator for alarming the ful...

Page 9: ...umping open the drain valve to drain the exhaust 0 0kPa to close drain valve intermittent steam generator to pump water to generate steam when the pot the temperature reaches the set value the end of...

Page 10: ...e provide accessories to take devices Must use handheld to take things prevent burns 8 Emergency Condition During the working time the Steam Sterilizer will automatically alarming then release pressur...

Page 11: ...to the right MAX position 7 E7 long du Fault of temperature senor on steam generator Inspecting the temperature sensor on steam generator 8 E8 long du Fail on pressure rising Inspect water flowing ro...

Page 12: ...ommendation by the supplier 9 5 The replacement work for sealing ring Tool One flat screwdriver No sharp edge at the head 1 One hand catches the lip of the sealing ring the other hand insert the screw...

Page 13: ...al components must cut off power supply first And only the manufacturer or professional person designated by the manufacturer can operate 10 Transportation and Storage 10 1 The preparatory work before...

Page 14: ...e door of the Steam Sterilizer in case of scald when open the door 9 Shut off the power before discharging installing the sealing ring and the work shall be carried out after further sufficient coolin...

Page 15: ...13 Appendix 1 Steam loop Work Description...

Page 16: ...14 2 Circuit diagram of sterilizer...

Page 17: ...playing Ed B D test when put the test material onto the tray inside the chamber firstly turn on the power by the green switch close the door keeping press the UNWRAPPED button around 10s When the boar...

Reviews: