3
ERR LED (red)
Status
Description
Off
Off or operating
Flashes once
Error counter has reached or exceeded preset value
Flickering
Automatically detecting baud rate
Flashes twice
NMT slave or NMT master is detecting Heartbeat event
(Heartbeat consumer)
On
Bus off
Connections
Interface Connector Pin Configuration
See "2-1. Rear Panel Interface Connector".
Cable Length
The correlation between baud rate and total distance is as follows:
Baud rate
Total distance
10 kbps
Up to 1,000 m
20 kbps
Up to 1,000 m
50 kbps
Up to 1,000 m
100 kbps
Up to 600 m
125 kbps
Up to 500 m
250 kbps
Up to 250 m
500 kbps
Up to 100 m
800 kbps
Up to 50 m
1 Mbps
Up to 25 m
Wiring precautions
Always turn off power before connecting or disconnecting the equipment.
Do not turn on the AC power until all installation work is complete. There is no power ON/OFF switch on
the unit itself.
Keep cables connected to the unit as far away as possible from sources of noise such as power lines or
control interfaces.
Always ground the equipment. Use Type D single grounding. Do not share a ground with the power
supply system.
Turn on the terminal resistance on the CANopen connector farthest from the PLC.
Use a dedicated CANopen cable as the connecting cable.
Select a CiA DR-303-1 compliant connecting cable.