14
3-3-5.
Actual Torque Setting
Set actual torque acting on the torque transducer (or torque transducer output).
When calibrating by entering numerical values, set the same value as set in "3-3-4 Torque Setting".
* Enter value (3) on the calibration certificate sheet. (No decimal point)
For example, if the maximum indication is 100.00 N·m, set to [10000].
Use the
keys to change the torque on
the lower display to the required value.
: Alters the value of the digit to be altered.
: Selects the digit to be altered.
: Cancels the setting and returns to
measurement
mode.
: Registers the value displayed and
proceeds to the next step.
Press the
key after setting.
Actual torque value
or torque meter output
To zero calibration
New torque value