3-3. Calibration Procedure
3-3-1.
Calibration in Measurement Mode
Step 1
Step 2
Step 3
Step 4
Step 5
Step 6
Step 7
Step 8
Step 9
Connect the torque transducer to the equipment.
Turn on and wait for approximately 10 minutes for the torque
transducer and equipment to stabilize.
Select calibration mode.
Select the minimum torque units.
Select the minimum torque units.
Set the maximum torque value that can be measured by the
torque transducer.
Calibrate the zero point with the torque transducer at the initial
torque state or by setting the zero point torque for the torque
transducer.
Calibrate the span point with the span torque acting on the torque
transducer or by setting the span point torque for the torque
transducer.
Calibration is complete. Perform steps 10 onward as described
below where necessary.
Steps 4 to 9 must be carried out in sequence. The calibration results will not be updated if the procedure is not
completed up to step 9.
The A/Z function is canceled once calibration is completed (up to step 9).
The A/Z function is canceled if step 10, 11, or 12 is performed.
For details of steps 11 to 15, refer to Instruction Manual EN294-1596*.
Section
3-3-3
Minimum scale
division setting
Section
3-3-4
Torque setting
Step 11
Step 12
Step 13
Step 14
Section
3-3-5
Actual torque setting
Section
3-3-7
Span calibration
Section
3-3-6
Zero calibration
Step 15
After turning on power
Calibration mode
selection
End of calibration
Decimal place
setting
Units setting
Symmetry
correction
function
Linearization
correction
function
Polarity change
function
Step 10
Section
3-4
Zero/span fine
adjustment
function
Torque transducer
connection