9
Maintenance/repairs/cleaning
9.1
Repairs
9.2
Cleaning
9.2.1
Cleaning instructions
The device should be cleaned at regular intervals.
Disconnect the device from the supply voltage before cleaning.
Remove all traces of corrosive substances on a regular basis.
If the scale is located in a dry environment, then wipe the weighing platform using a
damp cloth.
Devices with IP rating IP65 or higher can also be rinsed with a water jet directed at the
load plate from above.
Do not allow liquid to enter devices with an IP43 degree of protection.
Condensation may form in the device if it is cleaned with water that is too hot or too cold,
due to temperature differences. Condensation may cause the device to malfunction.
9.2.2
Cleaning agents
NOTICE
Some cleaning agents may not be compatible with the device material.
Do not use cleaning agents for stainless steel parts that contain chlorine, alkalines,
acetic acid, hydrochloric acid, sulfuric acid, or citric acid.
The use of cleaning sponges made of steel wool is prohibited (e.g. S.O.S pads).
Use a damp cloth or sponge to clean stainless steel parts on the scale (if present).
You can use any commercially available household cleaning agent that is suitable
for use on stainless steel.
Suitable cleaning agents for the painted version:
-
standard soft brooms and hand brushes,
-
compressed air,
-
damp cloths, detergents, all-purpose cleaners for adhering dirt.
Cleaning agents suitable for the stainless steel version:
-
Chlorine cleaner with max. 6 % chlorine content
-
All
‑
purpose cleaner
-
Industrial cleaners
Floor scale Puro®
9 Maintenance/repairs/cleaning
EN-48