The adjustment weight is displayed once adjustment is inished.
GMP-Protokoll wird ausgedruckt.
--------------------
Typ CAW1P1-3DC-L
Ser.no. 12345678
Vers. C1 111.031112
BVers. 01-62-02
Ser.no. A 12345678
--------------------
Internal calibration
Diff. + 0.2 g
Internal adjustment
Diff. + 0.0 g
--------------------
Name:
--------------------
7.4 SQmin-Funktion
To display the allowable minimum sample quantity "SQmin" (sample quantity minimum)
in accordance with the United States Pharmacopoeia (USP). According to USP guidelines,
the uncertainty of measurement may not exceed 0.1 % of the sample quantity when
The diference between the measured value and the true weight of the sample
will be displayed with plus/minus signs.
A printout will be generated if the process is cancelled using the
key.
Internal calibration
Diff. + 0.2 g
3. Activatecalibration/adjustment with
key. (press the
key to cancel).
7 Operating
Combics 1 Series Complete Scales Models CAW1P, CAW1S, CAS1
Minebea Intec