*: Blank space
A: Characters of the display (max. 7 characters and decimal point)
E: Characters for the unit (1 … 3 letters, followed by 2 … 0 blank spaces)
CR: Carriage Return
LF: Line Feed
Special operation
Position
1
2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
*
*
*
*
*
*
*
– – *
*
*
*
*
CR LF
or
*
*
*
*
*
*
*
H *
*
*
*
*
*
CR LF
or
*
*
*
*
*
*
*
H H *
*
*
*
*
CR LF
or
*
*
*
*
*
*
*
L
*
*
*
*
*
*
CR LF
or
*
*
*
*
*
*
*
L
L
*
*
*
*
*
CR LF
or
*
*
*
*
*
*
*
C *
*
*
*
*
*
CR LF
*: Blank space
– –: Output weight
H: Overload
HH: Checkweigher overload
L: Underload
LL: Checkweigher underload
C: Calibration
Error message
Position
1
2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
*
*
*
E r
r
*
*
# # *
*
*
*
CR LF
or
*
*
*
E r
r
*
*
# # # *
*
*
CR LF
*: Blank space
#: Number (2
‑
or 3
‑
igure error number)
Example
:
Output of the weight value +1255.7 g
Position
1
2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
+ *
*
*
1
2 5 5 .
7
*
g *
*
CR LF
Position 1: Algebraic sign +, - or blank space
Position 2: Blank space
Position 3 … 10: Weight value with decimal point. Leading zeros are displayed as blank
spaces.
Position 11: Blank space
Position 12 … 14: Character for measurement unit, blank space or character
!
as symbol
Position 15: Carriage Return
Position 16: Line Feed
Combics 1 Series Complete Scales Models CAW1P, CAW1S, CAS1
8 Data interfaces
EN-104
Minebea Intec