32
Operating Instructions Combics 3 Filling
Data Interfaces
GMP-Compliant Printouts
When this function, the printout is supplemented with a GMP header and a
GMP-footer (GMP: Good Manufacturing Practice).
If the GMP-compliant printout is activated, the
A
symbol remains displayed until
the GMP footer is printed.
Setting: Setup menu under
“
Device parameters: Printout: ISO/GLP/GMP printout
”
You can choose from the following settings:
– GMP-compliant printout off (factory setting)
– GMP-compliant printout for 1 application result
– GMP-compliant printout for multiple application results
The GMP header is included from the first printout generated subsequent to the
activation of the GMP printout.
The GMP footer is printed after multiple measurement results when the
p
key is pressed and held. In this case, the
A
symbol remains visible until the
GMP-footer is printed.
If the scale is changed during a GMP printout for multiple measurement results,
the GMP footer for the previously used scale is printed upon pressing of the
n
key. The GMP header for the other platform is included on the next printout
generated.
A GMP-compliant printout is generated automatically at the conclusion of
calibration/adjustment and linearization routines, as well as when you set or
delete a preload.
If you use a label printer for GMP-compliant printouts, you may find that a single
label is not long enough for the data printed. If this is the case, you can activate
the automatic form feed after each printout of a GMP header and measurement
results. Sample GMP headers and footers are provided (see “Sample Printouts” in
the application manual).