3.
In the [
Calibrate CO
2
] menu, select [
Main Menu
] → [
Maintenance >>
] → [
User
Maintenance >>
] → enter the required password → [
Module Maintenance >>
] →
[
Maintain CO
2
>>
] → [
Calibrate CO
2
>>
].
4.
In the [
Calibrate CO
2
] menu, select [
Zero
].
5.
After the zero calibration is finished successfully, connect the equipment as follows:
6.
Open the relief valve, and adjust it until the flowmeter has a stable reading between 10
ml/min and 50 ml/min.
7.
In the text box on the upper right corner of the [
Calibrate CO
2
] menu, enter the vented
gas concentration.
8.
In the [
Calibrate CO
2
] menu, the measured CO
2
concentration is displayed. After the
measured CO
2
concentration becomes stable, select [
Calibrate
] to calibrate the CO
2
module.
9.
If the calibration is finished successfully, the message [
Calibration Completed!
] is
displayed. If the calibration failed, the message [
Calibration Failed!
] is displayed. In
this case, perform another calibration.
5.5 Nurse Call Relay Performance Test
Tools required:
Oscilloscope
1.
Connect the nurse call cable to the multifunction output connector of the patient
monitor.
2.
Enter Demo mode. Then, select [
Main Menu
] → [
Maintenance >>
] → [
User
Maintenance >>
] → enter the required password → [
Others >>
].
Monitor
Gas cylinder
Flowmeter
Reducing
valve
T-shape connector
Sampling line