Quick Reference Guide

Summary of Contents for Passport 12

Page 1: ...Quick Reference Guide ...

Page 2: ...e is not intended as a replacement to the Operating Instructions Prior to operating equipment the user must be familiar with the Operating Instructions Manual contents This document is a guideline only to be used as an aid to comprehensive Inservice Training ...

Page 3: ... And Set Pacemaker Rate 6 Initiate Relearn Arrhythmia Option Required 7 Initiate ST Analysis ST Option Required 7 NIBP 7 Initiate A Non Invasive Blood Pressure Measurement NIBP 7 Set A Non Invasive Blood Pressure Measurement Interval 8 Invasive Blood Pressures Optional 9 Adjust Invasive Blood Pressure Label 9 Zero Invasive Blood Pressure 9 Alarms 10 Adjust Alarm Limits 10 Adjust Alarm Responses 11...

Page 4: ...ends 20 Arrhythmia And Alarm Events 20 Full Disclosure 21 Volumes 22 Key Volume 22 Beat Tone 23 Alternate Operating Modes 23 Night Mode 23 Privacy Mode 24 Date And Time 25 Parameter Colors 26 Load Default Preset Settings 26 Lock Unlocking Touchscreen Display 27 Cleaning Disinfecting 28 Cleaning the Main Unit Module 28 Disinfecting the Main Unit Module 30 Cleaning and Disinfecting the Accessories 3...

Page 5: ...atient setup menu Trigger a manual event Have a split screen view of minitrend Enter the volume setup menu Default configurations Start cardiac output procedure Perform calculations Have a split screen view of another patient s conditions Have a split screen view of OxyCRG trends Enter the full screen 7 lead ECG screen Enter the parameters menu Start NIBP STAT measurement Enter the unit setup menu...

Page 6: ...g onscreen keyboard f Select to accept the information g Select OK to return to main display DPM Central Station Interface Entering demographic information on the DPM Central Station automatically enters it at the bedside monitor Entering demographic information on the bedside monitor automatically enters it on the DPM Central Admit button selection is required for patient information to be displa...

Page 7: ...or First Name and or MRN f Enter patient s information using onscreen keyboard g Select Query h Select patient and select Import i Select OK to return to main display Optional Installation Required Panorama Central Station Interface Entering demographic information on the Panorama Central Station or WorkStation automatically enters it on the bedside monitor Entering demographic information on the ...

Page 8: ...ng Patient Setup quick key or Main Menu quick key b Select Discharge Patient c Select OK to confirm Note After the patient is discharged default preconfigured settings are restored and the monitor can be placed in Standby DPM Central Station Interface Discharging a patient on the Passport 8 or Passport 12 bedside monitor will automatically discharge the patient on DPM Central Station Discharging t...

Page 9: ...l Station in standby Select any key on the Passport 8 or Passport 12 bedside monitor to remove patient from Standby Selecting Standby in the patient s drop down menu will remove patient from standby Panorama Central Station Interface Selecting Standby on the Passport 8 or Passport 12 bedside monitor automatically places the Panorama Central Station WorkStation and ViewStation in standby Selecting ...

Page 10: ...Interface Selecting Real Time Print on the Passport 8 or Passport 12 bedside monitor will print a 10 second Real Time Report on the DPM Central Station The waveforms printed are determined by the Passport 8 or Passport 12 bedside monitor Panorama Central Station Interface Selecting Real Time Print on the Passport 8 or Passport 12 bedside monitor will print a 10 second All Strips Report on the Pano...

Page 11: ... Central Station WorkStation and ViewStation are independent Enable Pacemaker Detection And Set Pacemaker Rate The Passport 8 or Passport 12 bedside monitors allow the clinician to enable pacemaker detection The symbol is displayed when the pacemaker detection is enabled To enable the pacemaker detection on the Passport 8 or Passport 12 bedside monitor a Select ECG waveform or ECG numeric tile to ...

Page 12: ...he main display Initiate ST Analysis ST Option Required To initiate ST analysis on the Passport 8 or Passport 12 bedside monitor a Select ECG Setup by selecting ECG waveform HR numeric tile or from Parameters menu or Main Menu quick keys b Select ST Analysis c Select ON for ST analysis value d Select Main Menu quick key or to return to the main display NIBP Initiate A Non Invasive Blood Pressure M...

Page 13: ...elect Main Menu quick key or to return to the main display DPM Central Station Interface Selecting Interval in the NIBP menu ViewBed Tab on the DPM Central Station allows the clinician to set an automatic interval on the Passport 8 or Passport 12 bedside monitor To stop the current NIBP measurements on a patient on the Passport 8 or Passport 12 bedside monitor a Select NIBP on the front panel of t...

Page 14: ...BP Module as needed b Select the Invasive Blood Pressure ART CVP etc Setup by selecting IBP waveform tile or from Parameters menu or Main Menu quick keys c Select the invasive blood pressure ART CVP etc from the Parameters menu in Main Menu quick key to enter the Invasive blood pressure s Setup menu d Select label and adjust the label using the use the up and down arrow keys e Select Main Menu qui...

Page 15: ... or ST tab to change ST limits c Scroll and select the parameter d Select High or Low limit value and adjust using the up and down arrows e Select Off in the limit value area to disable the parameter s alarm limit s f Select Main Menu quick key or to return to the main display DPM Central Station Interface Changing alarm limits on the DPM Central Station automatically change alarm limits on the Pa...

Page 16: ...arm settings to designated laser printer Default to restore preset settings d Select Main Menu quick key or to return to the main display Or a Select Alarm Setup from quick key or Main Menu quick key b Select Parameters tab to change parameters or ST tab to change ST c Scroll and select parameter d Select On Off to enable or disable the alarm s limits or responses Level to adjust the audio and vis...

Page 17: ...the Passport 8 or Passport 12 bedside monitor a Select Alarm Setup from quick key or Main Menu quick key b Select Arrh Analysis tab c Scroll and select individual arrhythmia d Select Alarm to enable or disable arrhythmia alarm notification Alm Lev to adjust the audio and visual alarm indicators Level options include High Red Medium Yellow and Low Yellow and Message audio absent Alm Rec to activate...

Page 18: ...ect Alarm Setup from quick key or Main Menu quick key b Select Arrh Threshold tab c Scroll and select individual arrhythmia threshold d Adjust value using the up and down arrows e Select Main Menu quick key or to return to the main display RESET ALARMS To silence an alarm for the duration of the alarm on the Passport 8 or Passport 12 bedside monitors select Alarm Reset key on the front panel of be...

Page 19: ... Panorama Central Station Interface Selecting Alarm Pause on the Passport 8 or Passport 12 bedside monitor will silence active alarms on the Panorama Central Station WorkStation and ViewStation for a preconfigured amount of time Selecting Mute on the Panorama Central Station will not silence alarms on the Passport 8 or Passport 12 bedside monitor TRANSFER The Passport 8 or Passport 12 bedside moni...

Page 20: ...ort 8 or Passport 12 bedside monitor Panorama Central Station Interface Transferring patient data from one Passport 8 or Passport 12 bedside monitor to another Passport 8 or Passport 12 bedside monitor will not transfer data within the Panorama Central Station Transferring data within a Panorama Central Station or to a different Panorama Central Station may affect the Passport 8 or Passport 12 bed...

Page 21: ...eded h Select measurement to include exclude in the Cardiac Output average a in the box indicates cardiac output measurement is selected i Select 1 Cancel to stop the current measurement 2 Record to record the most recent cardiac output measurement and numeric data j Select Main Menu quick key or to return to the main display Or j Select Calc to access the Hemodynamic Calculations menu Note Blood ...

Page 22: ...nter the PAWP measurement g Select 1 Next Measurement to start a new measurement as needed and repeat previous steps 2 Exit to return to normal screen h Select Calc to access the Hemodynamic Calculations menu Note PA alarms are inactive when a PAWP measurement is initiated Note Follow manufacturer s suggested procedures and hospital policy for PAWP balloon inflation Initiate Hemodynamic Calculatio...

Page 23: ...tion can be printed and stored on the DPM Central Station SCREEN DISPLAY The Passport 8 or Passport 12 bedside monitor allows the clinician to adjust the main display based on patient acuity To enter the screen selection menu and adjust the main screen s layout on the Passport 8 or Passport 12 bedside monitor a Select Screens quick key on main display Or a Select Main Menu quick keys 1 Select Scre...

Page 24: ...ptional 12 leads of ECG with numeric parameters on the main display for Passport 12 bedside monitor only PAWP Screen Optional Split screen with PAWP measurement window on the lower portion of the main display with real time waveforms and numeric parameters C O Screen Optional Split screen with C O measurement window on the upper portion of the main display with real time waveforms and numeric para...

Page 25: ...itor to the integral recorder a Select Review quick key or from Main Menu quick key b Select Tabular Trends tab c Select Record and confirm specified time period d Select Record to print to the integrated recorder e Select Print to print to the independent printer f Select Main Menu quick key or to return to the main display DPM Central Station Interface Trend storage display interval and printing...

Page 26: ... Passport 12 bedside monitor DPM Central Station Interface Event storage and printing length on the Passport 8 or Passport 12 bedside monitor and DPM Central Station are independent Panorama Central Station Interface Event storage and printing length on the Passport 8 or Passport 12 bedside monitor and Panorama Central Station are independent Full Disclosure The Passport 8 or Passport 12 bedside m...

Page 27: ... Panorama Central Station are independent VOLUMES The alarm volume key volume and QRS volume can be adjusted on the Passport 8 or Passport 12 bedside monitors To adjust the alarm volume on the Passport 8 or Passport 12 bedside monitor a Select Volume quick key or select Others Tab in Alarm Setup quick key b Select Alm Vol c Select volume ranging from X 10 0 Off 10 Max d Select Main Menu quick key ...

Page 28: ... quick key or to return to the main display Note The QRS tone is derived from either the HR or PR depending on which is currently selected as the alarm source When monitoring SPO2 there is a variable pitch tone which changes as the patient s saturation level changes The pitch of the tone rises as the saturation level increases and falls as the saturation level decreases Alternate Operating Modes T...

Page 29: ...e To suspend alarm notification and turn off the display on the Passport 8 or Passport 12 bedside monitor when admitted to the DPM Central Station a Select Privacy Mode quick key Or a Select Main Menu quick keys 3 Select Screen Setup 4 Select Privacy Mode b Select Yes To remove a patient from Privacy Mode and resume monitoring on the Passport 8 or Passport 12 bedside monitor select any key on the ...

Page 30: ...t Time Format to adjust the time format 12h or 24h g Select Main Menu quick key or to return to the main display DPM Central Station Interface Adjusting the date and time on the DPM Central Station will adjust the date and time on the Passport 8 or Passport 12 bedside monitor The date and time cannot be adjusted on the Passport 8 or Passport 12 bedside monitor when interfacing with the DPM Central...

Page 31: ...r Passport 12 bedside monitor and DPM Central Station are independent Panorama Central Station Interface Parameter color on the Passport 8 or Passport 12 bedside monitor Panorama Central Station ViewStation and WorkStation are independent LOAD DEFAULT PRESET SETTINGS The Passport 8 or Passport 12 bedside monitors settings are automatically restored either when the patient size is adjusted or when ...

Page 32: ...bility to lock and unlock the touchscreen display To lock the touchscreen display on the Passport 8 or Passport 12 bedside monitor press and hold Main Menu quick key for 2 3 seconds until appears To unlock the touchscreen display on the Passport 8 or Passport 12 bedside monitor press and hold Lock quick key for 2 3 seconds until Main Menu quick key appears ...

Page 33: ...roxide Hydrogen peroxide 3 Isopropanol Isopropanol 70 1 Propanol 1 Propanol 50 CIDEX OPA Solution Ortho Phthalaldehyde 0 55 Metrex CaviCide1 Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0 28 Isopropanol 17 2 Virex II 256 Didecyl dimethyl ammonium chloride 8 704 n Alkydimethyl benzyl ammonium chloride 8 190 Virex TB n Alkyl dimethyl benzyl ammonium chlorides 0 105 n Alkyl dimethyl...

Page 34: ... Healthcare Hydrogen Peroxide Cleaner Disinfectant Wipes Hydrogen Peroxide 1 4 Diversey Oxivir TB Wipes Hydrogen Peroxide 0 5 Metrex CaviWipes Diisobutylphenoxyethoxyethyl dimethyl benzyl ammonium chloride 0 28 Isopropanol 17 2 PDI Sani Cloth AF3 Germicidal Disposable Wipe n Alkyl dimethyl ethylbenzyl ammonium chlorides 0 14 n Alkyl dimethyl benzyl ammonium chlorides 0 14 PDI Sani Cloth Bleach Ger...

Page 35: ... equipment without the symbol only the cleaning and disinfecting agents marked with are available for use To clean the equipment a Shut down the monitor and disconnect it from AC power b Clean the display screen with wipes or a soft cloth moistened with one of the cleaning agents listed in the table above c Clean the exterior surface of the equipment or module with wipes or a soft cloth moistened ...

Page 36: ...fecting of accessories please refer to the manufacturer s packaging or the Passport 8 Passport 12 Patient Monitor Operator s Manual STERILIZATION Sterilization is not recommended for this equipment related products accessories or supplies unless otherwise indicated in the Instructions for Use that accompany the products accessories or supplies ...

Page 37: ...Mindray DS USA Inc 800 MacArthur Blvd Mahwah NJ 07430 Tel 800 288 2121 or 201 995 8000 www mindray com 0 4 6 0 0 5 8 4 7 0 0 3 0 ...