the patient tile at the Panorama Central Station and the
SpO
2
tile is removed.
The heart rate color in the patient tile has changed while
the SpO
2
module is being used. What is the reason?
A patient’s heart rate is obtained (sourced) from the ECG or
the SpO
2
when using a Panorama Telepack. If an ECG lead
off condition occurs a “Lead OFF” message is displayed
in the patient tile on the Panorama Central Station. If the
heart rate is unobtainable from the ECG, the heart rate is
automatically obtained from the SpO
2
. The heart rate color
will change to the SAME color as the SpO
2
parameter to
indicate this shift in the heart rate source.
Note: The arrhythmia package is temporarily disabled
while the heart rate is being obtained from the SpO
2
parameter.
Are there any other reminders or special considerations
for the Panorama Telepack SpO
2
module?
a. Remove the SpO
2
module prior to defibrillating a
patient
b. Remove the SpO
2
module prior to entering the MRI
environment
c. Insert SpO
2
module cleaning plug prior to cleaning
the SpO
2
module to maintain its water resistance
rating