b. ECG lead wires disconnected
• Attach ECG lead wires sets to Panorama Telepack
c. SpO
2
module not completely connected
• Turn off the Panorama Telepack by rotating
battery door counterclockwise
• Disable the SpO
2
module by pressing the
release button
• Align SpO
2
module with Panorama Telepack
and slide module onto telemetry pack until an
audible “click” is heard
• Ensure ECG lead wires set is connected to
Panorama telepack
• Turn Panorama Telepack on by rotating battery
door clockwise
d. SpO
2
patient cable and sensor connection issue
• Verify connection on bottom of SpO
2
module
• Verify connection of sensor with SpO
2
patient
cable (disposable probe)
• Verify SpO
2
sensor is attached
• Ensure red light is illuminated on sensor
i. Replace patient cable and/or sensor if not
illuminated
ii. Replace batteries in Panorama Telepack
e. Display Tab is not configured to display SpO
2
• Select
PATIENT SETUP
menu button
8
P/N: 0077-00-0003-01 Rev B
Panorama
®
Telepack-608 SpO
2
Quick Reference Guide