Structure and Assembly/Disassembly of the Machine
4.7Other Faults
Table 5-2 Other faults list
No
Fault
Cause
Remark
1
The fan doesn’t
work.
+12V power is abnormal
◆
The cable’s fault
◆
The fan’s fault
◆
Supply the
+
12V power
directly by the power
board, check the +12V
voltage in the power
board.
2
USB fault
The cable’s fault
◆
The main board is abnormal
◆
The software doesn’t
◆
support U disk or the printer.
Replace USB interface
◆
cable
Check the main board
◆
Use U disk or the
◆
printer of specified
model
3
Video output fault
Video cable’s fault
◆
(
inside
the main board
)
The main board is abnormal
◆
Check
the
◆
corresponding
video
cable
Check the main board
◆
4
Network fault
The main board is abnormal
◆
First verify that the
◆
external network is
connected correctly.
DP-3300/DP-3200
Service Manual
(
V1.1
)
5-5