Structure and Assembly/Disassembly of the Machine
system
Boot bitmap
The image displayed when the
system starts up.
BMP
The system
resets
automatically
Factory
default setup
the backup of factory DTA data
in flash
DTA
The system
resets
automatically
Chinese entry
Chinese entry data
CIM
The system
resets
automatically
Transducer’s
data
The data of the transducers
supported
PRB
The system
resets
automatically
The system software is taken as an example to describe the upgrading procedure:
Put a U disk with UPG upgrading files into a USB interface.
→Click “system software”, a file dialogue box pops up and opens:
Load Fi l e
Hel p
Dri ver: C
:
Path:
Fi l es: DP6600 . UPG
Total System Update Fi l es 1
:
C:
OK
Cancel
Name
Type
Modi fy ti me
Pati ent Name
Pati ent I D
2
3
4
1
UPG
DP6600
04- 11- 20 14: 20
Figure 4-45 The screen for the software upgrading
→Select the file to be upgraded in the dialogue box. The meaning of each item is
described as follows:
1. Drive path
2. Directory path
3. Name of the selected file
4. File list
DP-3300/DP-3200
Service Manual
(
V1.1
)
4-3
Summary of Contents for DP-3200
Page 1: ...DP 3300 DP 3200 Digital Ultrasonic Diagnostic Imaging System Service Manual...
Page 2: ......
Page 11: ...DP 3300 DP 3200 Service Manual V1 1 I...
Page 12: ......
Page 16: ......
Page 20: ......
Page 76: ......
Page 81: ...P N 2302 20 34499 V 1 1...