46
MW806
Portable Meter
Note: The USB flash drive can be safely removed if the USB icon
is not displayed. Do not remove the USB drive during export.
Overwriting existing data:
When the LCD displays “EXPORT” with ACCEPT tag and lot number
blinking (USB icon displayed), an identically named lot exists on
the USB.
1. Press GLP/ACCEPT to continue. “OVERWRITE” is displayed
with ACCEPT tag blinking.
2. Press GLP/ACCEPT to confirm. Not confirming exits the export.
Display returns to lot selection screen.
Data Management Warnings
“NO MANUAL / LOGS” No manual records saved.
Nothing to display.
“NO STABILITy / LOGS” No stability records saved.
Nothing to display.
"OVR" with lot ###
(blinking)
Identically named lots on USB.
Select overwrite option.
"NO MEMSTICK"
USB drive is not detected.
Data cannot be transferred.
Insert or check the USB flash drive.
"BATTERy LOW"
(blinking)
When low battery, export is not
executed. Replace the battery.
Logged Data Warnings in CSV file
°
C !
Probe used beyond its operation
specifications. Data not reliable.
°
C !!
Meter in MTC mode.
"Log end ‑ Probe
disconnected"
Data logged with the probe disconnected
or damaged.