26
MW806
Portable Meter
Calibrating with Custom Buffers
Custom buffer value needs to be configured in pH mode Setup.
Temperature compensation of custom buffers is set to the value
of 25 °C.
y
Press RANGE/
. The custom buffer value is displayed blinking
on the third LCD line.
y
Use the
keys to modify the value based on temperature
reading. The buffer value is updated after 5 seconds.
ºC
ATC
CB1
10.01
9.18
7.01
pH
CAL
25.0
1.7 5 P H
1.49
W A I T
ACCEPT
ºC
ATC
CB1
10.01
9.18
7.01
pH
CAL
25.0
1.7 5 P H
1.71
Note: When using custom buffers, CB1 and CB2 tags are
displayed. If only one custom buffer is used, CB1 is displayed
together with its value.
Clear Calibration
1. Press CAL/EDIT to enter Calibration mode.
2. Press LOG/CLEAR.
ACCEPT tag is displayed blinking and “CLEAR CAL“ message
is displayed on the third LCD line.
3. Press GLP/ACCEPT to confirm.
“PLEASE WAIT“ message is displayed, followed by the
“NO CAL” confirmation screen.
ACCEPT
ºC
ATC
7.01
pH
CAL
28.0
C A L
7.02
C L E A R
ºC
ATC
CAL
pH
MEASURE
28.0
7.02
N O C A L