44
MW806
Portable Meter
ACCEPT
RCL
LOG
66
C A L L
PH
L O G R E
ACCEPT
RCL
LOG
L O G P H
PH
C L E A R
10.2.3. Exporting Data
PC Export
1. With the meter on, use the supplied micro USB cable to
connect to a PC.
2. Press SETUP then CAL/EDIT.
3. Use the
keys and select “EXPORT TO PC”.
The meter is detected as a removable drive. LCD displays the
PC icon.
4. Use a file manager to view or copy files on the meter.
ACCEPT
USB
SETUP
T O P C
E X P O R T
ACCEPT
USB
SETUP
M E T E R
L O G O N
When connected to a PC, to enable logging:
y
Press LOG/CLEAR. LCD displays “LOG ON METER” with ACCEPT
tag blinking.
y
Press GLP/ACCEPT. Meter disconnects from the PC and the PC
icon is no longer displayed.
y
To return to “EXPORT TO PC” mode, follow steps 2 and 3 above.
Exported data file details:
y
The CSV file (comma separated values) may be opened with a
text editor or spreadsheet application.
y
The CSV file encoding is Western Europe (ISO‑8859‑1).
y
Field separator may be set as comma or semicolon. See
Separator Type, GENERAL SETUP OPTIONS section.
Note: File prefix depends on measurement log storage locations:
“PHLOT###” for pH logs, and “ECLOT###” for EC and TDS logs.