10. LOGGING
Logging locations are measurement mode specific. pH logs are
saved under “PH”, CONDUCTIVITy and TDS logs under “EC”.
y
Press LOG/CLEAR to log a measurement.
y
Press RCL to access or export logged data.
ACCEPT
RCL
LOG
66
C A L L
PH
L O G R E
ACCEPT
RCL
LOG
21
C A L L
EC
L O G R E
For log types and stability criteria, see Log Type, GENERAL SETUP
OPTIONS and DATA MANAGEMENT sections.
10.1. TYPES OF LOGGING
Manual log on demand
y
Readings are logged each time LOG/CLEAR is pressed.
y
All manual readings are stored in a single lot (i.e. records
made on different days share the same lot).
Log on stability
y
Readings are logged each time LOG/CLEAR is pressed and
stability criteria is reached.
y
All stability readings are stored in a single lot (i.e. records
made on different days are logged in the same lot).
Interval logging
Note: An interval logging lot can hold up to 600 records. When
an interval logging session exceeds 600 records, another log
file is automatically generated.
y
Readings are logged continuously at a set time interval
(e.g. every 5 or 10 minutes).
y
Records are added to the lot until the session stops.
y
For each interval logging session, a new lot is created.
A complete set of GLP information including date, time, range
selection, temperature reading and calibration information is
stored with each log.