background image

3

Replace 

Probe

Clean

Probe

Acceptable Slope

Performance

Probe

Clean

Probe

Replace 

Probe

Setting Calibration Points

  Menu Button>pH>pH Calibration

Before selecting calibration point(s), you may want to review the 

**Last Calib./GLP** screen. Scroll down to the “Last Calib./GLP” option 

and press

 SELECT. (Menu Button>pH>pH Calibration>Last Calib./

GLP)  

The **Last Calib./GLP** screen will then give you the date and 

time of the last calibration; indicate 1, 2, or 3 point calibration; and 

give you the calibration points, as shown below. 

 

In accordance with GLP (Good Laboratory Practices), you can 

scroll down on this screen to view Acid Slope and Base Slope 

once calibration is complete. 
1) On **pH Calibration** screen, select 1, 2, or 3 point calibration. 
2)  Depending on your selection, the **1 Pt Calibration**, 

**2 Pt Calibration**, or **3 Pt Calibration** screen will appear. 

Each of these screens lists the calibration points - various 

combinations of 4.01, 7.01, and 10.01. Place the cursor beside 

your desired point or combination of points and press SELECT
As an example, if you select 2 Point Calibration, a new screen 

will appear listing 7.01 & 4.01 Calib and 7.01 & 10.01 Calib. If 

you SELECT 7.01 & 4.01 Calib, another screen will appear which 

will take you through the calibration process.

 

  You will be instructed to “Place probe in 7.01 calibration   

    solution. Press Select.”

 

  A “Please Wait...” message will appear on screen.

 

  Once the probe calibrates in the 7.01 solution, you will then  

    be instructed to “Place probe in 4.01 calibration solution.  

     

Press Select.”

 

  Once the probe has calibrated in the 4.01 solution,  

 

    calibration is complete.

Using Acid/Base Slopes 

to Assess pH Probe Performance

In addition to displaying date, time, number of calibration 

points and specific points, the **Last Calib./GLP** screen 

provides readings for Acid Slope and Base Slope. Slopes are 

given as percentages and tell you how closely your freshly 

calibrated pH probe is reading as compared with an “ideal” 

probe. 
A new probe can get very close to “ideal” probe readings, but 

with aging and coating becomes less accurate. The following 

graph shows the acceptable range and indicates what should 

be done if the reading falls outside this range. If your probe is 

good in one range (acid or base), but bad in the other, it should 

be replaced.

    90%         95%                              102%      105%

                                  

A 3-point calibration will always give you an acid and a 

base slope: however, when using 1- or 2-point calibration, 

depending on points chosen, you may get an acid OR a base 

slope reading. In this case, the other reading will show as a 

default 99.9.

pH Probe Diagnostic

  Menu Button>pH>pH Probe Diagnostic

pH probes have a limited lifetime and must be replaced 

periodically to maintain accurate readings. If you suspect 

incorrect readings, work your way to the pH Probe Diagnostic 

screen. Press SELECT. Follow instructions on the screen and 

the unit will run a diagnostic test, then a message will appear 

on the screen, such as “Probe or solution is bad!” Replace 

your solution, run diagnostic again, and if you get the same 

message, replace the probe. 

Understanding pH Alerts (Caution and Critical)

The DL100 is designed to efficiently monitor the pH value of 

a solution and generate caution and critical alerts as needed. 

The status lights located beside the screen will change color to 

signal these alerts and, if you have enabled Wi-Fi and subscribe 

to the interface portal, the monitor will send out your choice of 

emails or text messages, or both, to alert you. See “Setting up 

Wi-Fi” section for details on connecting to the interface portal.
A Caution is intended to alert you when the pH reading is 

getting too high or too low. A Critical Alert is to let you know 

that a reading has reached a critical level and you have a 

serious problem needing immediate attention.
It is your responsibility to set the pH alerts for your specific 

requirements. This is a balancing act and may require some 

experimentation on your part. You want to receive the alerts 

when needed, but you don’t want the DL100 to be sending out 

alerts too frequently. See the diagram below for a typical setup.  

5.60          6.30           7.00          7.70           8.40

Critical     Caution                     Caution     Critical

Red  |    Lime    |              Blue              |    Lime    |  Red

Below  Alert                                                 Above Alert

If the value for pH is within the Caution parameters you have set, 

the status lights on both sides of the screen will remain blue.
If the blue changes to lime, this signals that the value has 

reached the Caution level. If the reading continues to drift 

further from the target value and reaches the Critical level, the 

lime lights will turn red, indicating a critical condition.

** Last Calib./GLP **

O5/04/15   9:25

2 Point Calibration

7.01 & 4.01

          Acid Slope:  000.0

          Base Slope: 000.0

       >Exit

Summary of Contents for DL100

Page 1: ...1 DL100 BLACK BOX DATA LOGGING MONITOR pH Temperature Set Up Operation Manual ...

Page 2: ...asic Setup pH Data Log Setup Last Logged Wi Fi Settings Exit Wi FiSettings Wi FiEnable SecurityType NetworkName SSID Pre sharedKey ViewDeviceIP UploadInterval PortalSync Portal Service Exit Wi FiEnable Enable Disable Wi FiSettings Wi FiEnable SecurityType NetworkName SSID Pre sharedKey ViewDeviceIP UploadInterval PortalSync Portal Service Exit Wi FiSecurity Open WEP WPA Exit Wi FiSettings Wi FiEna...

Page 3: ...eens only four lines will be visible at a time To see more options scroll the list by using up down arrows on the keypad When the cursor appears beside the appropriate function or setting press SELECT THE KEYPAD Pressing the Menu button brings up a screen featuring the full menu At the bottom of all menu screens there will be an Exit option The Exit option when selected takes you back to the previ...

Page 4: ...inutes 00 00 using left right arrows picking the number you want to change then using up down arrows to select the appropriate number Note that this unit displays in military time 24 hour there is no am pm option that is 9AM is 09 00 and 2 05PM is 14 05 10 30PM is 22 30 9 Once complete press SELECT to enter the new time A screen will appear indicating that your change has been SAVED 10 Scroll to E...

Page 5: ...er it should be replaced 90 95 102 105 A 3 point calibration will always give you an acid and a base slope however when using 1 or 2 point calibration depending on points chosen you may get an acid OR a base slope reading In this case the other reading will show as a default 99 9 pH Probe Diagnostic Menu Button pH pH Probe Diagnostic pH probes have a limited lifetime and must be replaced periodica...

Page 6: ...nterval Start Time Please note that if you remove the SD card when the unit is ON the unit will remain operational However before re inserting the SD card disconnect the unit from its power source Once you have the SD card snapped back in place restore power to the unit If you re insert the SD card while the unit is operational you must still momentarily disconnect from power then reconnect allowi...

Page 7: ...e appropriate security type and press SELECT Menu Button Wi Fi Settings Network Name SSID Remember to use 2 4 GHz only 1 Work your way to the Network Name SSID screen and press SELECT 2 When the Network Name SSID screen opens fill in your Wi Fi name using the keypad arrows and press SELECT Menu Button Wi Fi Settings Pre shared Key Password 1 Work your way to the Pre shared Key screen and press SEL...

Page 8: ...ly International Plug Set No Converter Needed pH Electrode MA911B 2 Main Board Atmel SAM3X8E ARM Cortex M3 CPU Operating Voltage 3 3 volts Clock Speed 84 MHz clock S Ram 96 KBytes of SRAM two banks 64KB and 32KB Flash Memory 512 KB Data Retention 100 years at 25 C Station and Access Point Wi Fi 802 11 b g n Roles with Integrated Radio Baseband and MAC TCP IP Embedded TCP IP stack Circuit Board Pro...

Reviews: