background image

6

Manufactured/Assembled 

in the USA

Range

.001 pH to 14.000 pH

Resolution - User Select

Tenths (0.1 pH); Hundredths (0.01 pH); Thousandths (0.001) pH

Accuracy @ 77°F

0.02 pH

Setpoint Range

0.01 pH to 14.00 pH

Activation Parameter

User Select above or below activation setpoint

Visual Alerts

LED backlight blue = normal; lime = caution; red = critical

Probe Protection

On Master Board GLP chipset ( Ground Loop Preventer )

Circuit Board Protection

On Master Board RJ45 Backfeed Prevention Chipset

Operational Environment

15°F to 180°F  (-10°C to 85°C) 

Temperature Accuracy

0.5°F 

Power Supply

International Plug Set - No Converter Needed

pH Electrode

MA911B/2

Main Board

Atmel SAM3X8E ARM Cortex-M3 CPU

      Operating Voltage

3.3 volts

      Clock Speed

84 MHz clock

      S Ram

96 KBytes of SRAM (two banks: 64KB and 32KB)

      Flash Memory

512 KB

      Data Retention

100 years at 25° C

      Station and Access Point, Wi-Fi

802.11 b/g/n Roles with Integrated Radio, Baseband, and MAC

      TCP/IP

Embedded TCP/IP stack

       Circuit Board Protection

RJ45 Back Feed Protection with Short Circuit Protection

       GLP (Ground Loop Prevention)

 Dual Ground Loop Prevention Circuit and Chipset

       Input Power Supply

 7V to 12VDC Recommended --- International Plug Set - No Converter Needed

       Input Min. & Max.

 6V to 20VDC

       Logic Load Shifter

 Provides Higher LCD Screen Resolution and Fast Response Time

DL100 BLACK BOX SPECIFICATIONS

Summary of Contents for DL100

Page 1: ...1 DL100 BLACK BOX DATA LOGGING MONITOR pH Temperature Set Up Operation Manual ...

Page 2: ...asic Setup pH Data Log Setup Last Logged Wi Fi Settings Exit Wi FiSettings Wi FiEnable SecurityType NetworkName SSID Pre sharedKey ViewDeviceIP UploadInterval PortalSync Portal Service Exit Wi FiEnable Enable Disable Wi FiSettings Wi FiEnable SecurityType NetworkName SSID Pre sharedKey ViewDeviceIP UploadInterval PortalSync Portal Service Exit Wi FiSecurity Open WEP WPA Exit Wi FiSettings Wi FiEna...

Page 3: ...eens only four lines will be visible at a time To see more options scroll the list by using up down arrows on the keypad When the cursor appears beside the appropriate function or setting press SELECT THE KEYPAD Pressing the Menu button brings up a screen featuring the full menu At the bottom of all menu screens there will be an Exit option The Exit option when selected takes you back to the previ...

Page 4: ...inutes 00 00 using left right arrows picking the number you want to change then using up down arrows to select the appropriate number Note that this unit displays in military time 24 hour there is no am pm option that is 9AM is 09 00 and 2 05PM is 14 05 10 30PM is 22 30 9 Once complete press SELECT to enter the new time A screen will appear indicating that your change has been SAVED 10 Scroll to E...

Page 5: ...er it should be replaced 90 95 102 105 A 3 point calibration will always give you an acid and a base slope however when using 1 or 2 point calibration depending on points chosen you may get an acid OR a base slope reading In this case the other reading will show as a default 99 9 pH Probe Diagnostic Menu Button pH pH Probe Diagnostic pH probes have a limited lifetime and must be replaced periodica...

Page 6: ...nterval Start Time Please note that if you remove the SD card when the unit is ON the unit will remain operational However before re inserting the SD card disconnect the unit from its power source Once you have the SD card snapped back in place restore power to the unit If you re insert the SD card while the unit is operational you must still momentarily disconnect from power then reconnect allowi...

Page 7: ...e appropriate security type and press SELECT Menu Button Wi Fi Settings Network Name SSID Remember to use 2 4 GHz only 1 Work your way to the Network Name SSID screen and press SELECT 2 When the Network Name SSID screen opens fill in your Wi Fi name using the keypad arrows and press SELECT Menu Button Wi Fi Settings Pre shared Key Password 1 Work your way to the Pre shared Key screen and press SEL...

Page 8: ...ly International Plug Set No Converter Needed pH Electrode MA911B 2 Main Board Atmel SAM3X8E ARM Cortex M3 CPU Operating Voltage 3 3 volts Clock Speed 84 MHz clock S Ram 96 KBytes of SRAM two banks 64KB and 32KB Flash Memory 512 KB Data Retention 100 years at 25 C Station and Access Point Wi Fi 802 11 b g n Roles with Integrated Radio Baseband and MAC TCP IP Embedded TCP IP stack Circuit Board Pro...

Reviews: