If it is not possible to align the light sheet by lowering or raising the alignment tool, the actual position of
the horizontal focus does not match with the position displayed in the ImSpector software. Align the
central light sheet as well as possible before adjusting the horizontal focus.
Adjusting the horizontal focus
1
Select the
Advanced
tab on the bottom of the
Settings 1
window.
2
Move the slider
Sheetmotor Calibration
until only one line passes the pinhole. Ensure that the light sheet
focus is always in the center and the
Optics > sheet NA
is at the maximum. If necessary, readjust the z-
direction.
3
If only one line passes the pinhole in the center of the mask, activate the intensity profile.
4
Increase the intensity of this light sheet by switching between moving the alignment tool in z-direction and
moving the slider
Sheetmotor Calibration
. If not any further increase of the intensity is measurable, the
central light sheet is aligned.
5
Click
Save Settings
in the toolbar to fix the adjusted horizontal focus.
6
Click
Set as Zero
under
Current Position
to set a reference point of the alignment tool.
Summary of Contents for UltraMicroscope II
Page 1: ...UltraMicroscope II Imaging System User manual...
Page 14: ...1 IMPORTANT SAFETY INFORMATION 12...
Page 20: ...2 INTRODUCTION 18...
Page 30: ...5 PREPARING THE INSTRUMENT 28...
Page 38: ...6 LASER ALIGNMENT 36...
Page 46: ...7 SETTING UP AN EXPERIMENT 44...
Page 48: ...8 MAINTENANCE 46...
Page 54: ...10 TECHNICAL SUPPORT 52...
Page 57: ......