Laser alignment
6.1
Light sheet alignment
To generate high-contrast images with the best z-resolutions, all light sheets must be perfectly aligned to the
same z-plane. A set of adjustable precision screws on the left-hand and the right-hand side of the instrument
can be used to align the individual beams creating the light sheet. The Z screws enable an adjustment of the
light sheets in terms of the z-plane. Screw Z1 is used for aligning the central light sheet on the left-hand side in
the z-plane. Later the screw can be used for the fine adjustment of the z-shift between the left and right. The
screws Z2 and Z4 are used for aligning the outer light sheets. The x-ed out screws may not be adjusted. Always
start with the central light sheet before aligning the lateral light sheets.
Figure 6.1:
Side view of the instrument with the positions of the adjustable precision screws
Alignment tool
The alignment tool facilitates the adjustment of all light sheets. The mask carries an X with a pinhole in the
middle. All light sheets must pass the pinhole.
Figure 6.2:
Side view of the alignment tool
Summary of Contents for UltraMicroscope II
Page 1: ...UltraMicroscope II Imaging System User manual...
Page 14: ...1 IMPORTANT SAFETY INFORMATION 12...
Page 20: ...2 INTRODUCTION 18...
Page 30: ...5 PREPARING THE INSTRUMENT 28...
Page 38: ...6 LASER ALIGNMENT 36...
Page 46: ...7 SETTING UP AN EXPERIMENT 44...
Page 48: ...8 MAINTENANCE 46...
Page 54: ...10 TECHNICAL SUPPORT 52...
Page 57: ......