![Milltronics miniranger plus Instruction Manual Download Page 7](http://html1.mh-extra.com/html/milltronics/miniranger-plus/miniranger-plus_instruction-manual_1796962007.webp)
7ML19981CJ02
MiniRanger Plus – INSTRUCTION MANUAL
Page 3
Specifications
MiniRanger Plus
Fuse
• F1 2AG, FastBlow, 1.5 amp, 250 V
• F3 2 AG, SlowBlow, 1 amp, 250 V
• F4 Belling Lee, L754, 4000A HRC, ceramic type 100 mA, 250V
Environmental
• wall / rack / panel specific
Range
• solids applications: 7.5 m (25 ft.) (transducer dependent)
• liquid applications: 15 m (50 ft.) (transducer dependent)
Accuracy
• 0.25% of range or 6 mm (0.24"), whichever is greater
Resolution
• 2 mm
Memory
• program stored in non-volatile FLASH memory, for software upgrades via optional
Dolphin interface
• parameters stored in non-volatile EEPROM
Display
• wall / rack / panel specific
Outputs
• Transducer Drive
43.5 kHz, 250 V peak pulses of 1 ms max. duration at max.
repetition rate of 300 ms
• mA
optically isolated 0/4 - 20 mA
max. loading: 750
Ω
resolution: 0.1% of 20 mA
• Relays
2 alarm/pump control relays
1 form 'C' SPDT contact per relay, rated 5 A at 250 V AC
non-inductive
• Communication
wall / rack / panel specific
Approvals
• CE
1
, FM, CSA NRTL/C
1.
EMC performance available upon request.