Page 46
MiniRanger Plus – INSTRUCTION MANUAL
7ML19981CJ02
Example PMD Programming
If you do not have the characterization chart for the PMD then you will have to create the
chart yourself by measuring the head and flow at multiple points. It is much easier to use
the chart supplied by your PMD manufacturer.
All nine programmable breakpoints must be entered in ascending order (lowest to
highest) for OCM to work.
Run the PMD at different head levels and verify the OCM reading given on the
MiniRanger to double-check the programming. If this is not possible, double-check the
breakpoints for accuracy.
Note:
When the MiniRanger is in Program mode no measurements are taken and
the current total is reset to zero. This means that the totalizer relay (relay 2) will not
trip when the MiniRanger is in Program mode and the external totalizer will read
low.