Troubleshooting
54
LCD MESSAGE
WHAT IT MEANS
WHAT TO DO
ALARM, ALARM STOP
OR MAINTENANCE
MESSAGE?
DIST TEMPERATURE < 0
°
C
The measured temperature is < 0
°
C.
Contact Millipore SAS for assistance.
Programmable – can be
set to any type of
message (Maintenance
by Factory Default).
DIST TEMPERATURE > 50
°
C
The measured temperature is > 50
°
C.
Verify if the water is really this warm. Lower the Feedwater
Temperature.
Programmable – can be
set to any type of
message (Maintenance
by Factory Default).
SERVICE CONTRACT
REQUESTED
The Elix System is being maintained with a Service Contract. A
timer has indicated that a Service Contract visit is due.
Maintenance
(Yellow LED).
SYSTEM TEMPERATURE < 0
°
C
The measured temperature is < 0
°
C.
Contact Millipore SAS for assistance.
Programmable – can be
set to any type of
message (Maintenance
by Factory Default).
SYSTEM TEMPERATURE > 50
°
C The measured temperature is > 50
°
C.
Verify if the water is really this warm. Lower the Feedwater
Temperature.
Programmable – can be
set to any type of
message (Maintenance
by Factory Default).
TANK EMPTY
A SDS (a specific type of reservoir) is empty of water.
Verify that the reservoir is empty and is now filling up. Verify
that the Level Sensor from the SDS is connected to the Elix
System. This message will disappear when the water level is
10% full in the SDS.
Alarm (Red LED).
TOC > SETPOINT
The measured TOC is > the TOC Setpoint.
Check the value of the TOC Setpoint using the DATA Keypad
Button in PRODUCTION Mode.
Contact Millipore SAS.
Programmable – can be
set to any type of
message (Maintenance
by Factory Default).
TOC METER ERROR 0
A10 PCB E
2
Prom defective. Turn off/on the Elix System and
dispense water for several minutes. If the message continues,
then contact Millipore SAS.
Maintenance (Yellow
LED).
TOC METER ERROR 1
A10 PCB A/D converter defective. Turn off/on the Elix System and
dispense water for several minutes. If the message continues,
then contact Millipore SAS.
Maintenance (Yellow
LED).
TOC METER ERROR 2
A10 Thermistor defective. Turn off/on the Elix System and
dispense water for several minutes. If the message continues,
then contact Millipore SAS.
Maintenance (Yellow
LED).
TOC METER ERROR 3
Problem occurred with temperature compensation. Turn off/on
the Elix System and dispense water for several minutes. If the
message continues, then contact Millipore SAS.
Maintenance (Yellow
LED).
TOC METER ERROR 4
The water entering the A10 is < 4
°
C. Contact Millipore SAS.
Maintenance (Yellow
LED).
TOC METER ERROR 5
The water entering the A10 is > 41
°
C. Contact Millipore SAS.
Maintenance (Yellow
LED).