Maintenance
38
H
OW TO MANUALLY SANITISE THE
RO
C
ARTRIDGE
(
S
)
Make sure you have a Chlorine Tablet (see information above).
Make sure the Elix System has a Cleaning Port installed. It is installed as an accessory device and is not factory installed.
Place the Elix System into STANDBY Mode.
Locate the Cleaning Port Cap on the top of the Elix System. It is located towards the back top end of the Elix System
cabinet.
Unscrew the Cleaning Port Cap. Place the Chlorine Tablet into the Cleaning Port. Screw the cap back on.
Make sure the Cleaning Port Cap is tightened after putting in the Chlorine Tablet.
Additionally, make sure that the RO Reject Tubing is in a sink or drain. Make sure
it is fastened securely in place.
The Chlorine Table will form a strong chemical solution when mixed with water.
This solution can be dangerous for your eyes and your skin. Wear Eye Safety
Glasses. Use Laboratory Gloves and other appropriate safety equipment.
Press MAINTENANCE.
Press the Bottom Keypad Button.
Press Cl2 CLEAN for a few seconds. The Elix System will perform a Pressure Test for about 16 seconds.
The Elix System will go to RO CLEANING Mode. It will flush for 20 minutes.
!
HAZARD
S T A N D B Y
S E T U P
M A I N T E N A N C E
P R O D U C T I O N
M A I N T E N A N C E D A T A
E X C H P R O G A R D P A C K
E X C H A N G E P O L I S H E R
→
M A I N T E N A N C E
p H C L E A N
C l 2 C L E A N
→
M A I N T E N A N C E D A T A
E X C H P R O G A R D P A C K
E X C H A N G E P O L I S H E R
→
M A I N T E N A N C E
p H C L E A N
C l 2 C L E A N
→
T E S T I N G F E E D P R E S S .
P L E A S E W A I T . . .
R O C L E A N I N G
F L U S H I N G . . .
R . T I M E : 2 4 m i n
T E S T I N G F E E D P R E S S .
P L E A S E W A I T . . .
!
ATTENTION