Product Information 

Elix® 20, 35, 70 and 100 

2-2

 

S

YSTEM 

R

ECOVERY

 

W

HAT IS 

RO

 

R

ECOVERY

An RO Cartridge has 1 feedwater inlet and 2 water streams exiting it.  One exiting water stream is called the RO Permeate.  The 
other is called the RO Reject. 

The RO % Recovery is a comparison of the RO Permeate flow rate to the RO Feedwater flow rate.  It can be thought of as how 
much water you ‘recover’ from the RO as a useful product. 

It is calculated as: 

% Recovery = [(Permeate flow rate / Feedwater flow rate] * 100% 

Membrane % Recovery is defined as the recovery of a single RO Cartridge. 

System % Recovery is defined as the amount of water produced relative to the amount going into the entire Elix System. 

In most cases, the Membrane and System Recovery values will not be the same number. 

E

LIX 

S

YSTEM 

R

ECOVERY

 

An Elix System can be operated with a System Recovery between 30% to 70%.  This value will depend upon several factors 

including the chemical makeup of the Feedwater. 

System Recovery is defined above.  See “What is RO Recovery?” 

The Elix System Recovery is factory pre-set.  If desired, the Elix System Recovery can be modified.  Contact Millipore SAS for 
assistance. 

A System Recovery set too low will waste water.  A low recovery occurs when a lot of RO Reject Water is sent to drain relative to 
the RO Permeate flow rate (i.e. RO Recovery = 5%). 

A System Recovery set too high can possibly cause the RO Cartridges to become clogged.  A high recovery occurs when too little 
RO Reject Water is sent to drain relative to the RO Permeate flow rate (i.e. RO Recovery = 80%). 

Summary of Contents for Elix 35

Page 1: ...User Manual Elix 20 35 70 and 100 ...

Page 2: ......

Page 3: ... TOC CFU ml Eu ml when it leaves the water purification system provided that the systems are fed with water quality within specifications and properly maintained as required by the supplier We do not warrant these systems for any specific applications It is up to the end user to determine if the quality of the water produced by our systems matches his expectations fits with norms legal requirement...

Page 4: ... at a location that handles discarded electric or electronic equipment Proper disposal of equipment containing electric or electronic components will help to reduce pollution effects to the environment or to human health Proper recycling of these products helps in environmental preservation and helps to protect natural resources For more information about recycling of products containing electric ...

Page 5: ...oduct Water Quality 7 2 2 SYSTEM RECOVERY 8 What is RO Recovery 8 Elix System Recovery 8 2 3 SCHEMATIC OF MAIN COMPONENTS 9 Schematic of an Elix System 9 Guide to Schematic 9 RO Cartridge configuration for Elix Systems 10 2 4 DESCRIPTION OF INPUT AND OUTPUT CONNECTIONS 11 Ports 11 TANK LEVEL and RS 232 11 Grommets 12 2 5 DESCRIPTION OF MAIN COMPONENTS 13 Inlet Solenoid Valve 13 Flowmeter 13 Progar...

Page 6: ...NORMALLY DELIVERS PRODUCT WATER 23 SDS and Loops 23 Automatic Method 23 Manual Method 23 Alarm Message Displayed 24 3 2 ELIX WITH A POLISHER LOOP RESISTIVITY SENSOR OR A10 ADDED 25 3 3 HOW TO VIEW THE ELIX SYSTEM PERFORMANCE 26 3 4 HOW TO VIEW SERIAL NUMBERS AND MAINTENANCE INFORMATION FROM THE LCD 29 3 5 WHAT ARE THE DIFFERENT OPERATING MODES 31 CHAPTER 4 MAINTENANCE 33 4 1 SCHEDULED MAINTENANCE ...

Page 7: ...o an A10 Cleaning 45 How to see the remaining time of the A10 CLEANING 47 How to cancel an A10 CLEANING 47 4 9 HOW TO REPLACE THE A10 UV LAMP 48 4 10 HOW TO CALIBRATE THE A10 TOC MONITOR 48 CHAPTER 5 TROUBLESHOOTING 49 5 1 MAINTENANCE ALARM ALARM STOP AND CUSTOM MESSAGES 49 Types of messages 49 How messages are displayed and other information 49 5 2 AN ALARM MESSAGE OR MAINTENANCE MESSAGE IS DISPL...

Page 8: ......

Page 9: ... Catalogue Number The type of Elix System is indicated above the Keypad on the front of the Elix System Cabinet If this manual is not the correct one for your Water Purification System then please contact Millipore SAS INSTALLATION INFORMATION IMPORTANT INSTALLATION INSTRUCTIONS ARE NOT INCLUDED INSTALLATION OF THIS PRODUCT IS MEANT TO BE PERFORMED BY A QUALIFIED MILLIPORE SAS SERVICE REPRESENTATI...

Page 10: ...the next LCD Screen PRESS AND HOLD SYMBOL Pressing the Keypad Button next to this symbol for 2 3 seconds brings you to the next LCD Screen Pressing the Keypad Button next to this symbol brings you to the next LCD Screen within the same software branch Pressing the Keypad Button next to this symbol exits the currently viewed software branch It brings you back to the Root Menu Pressing the Keypad Bu...

Page 11: ...ing DATA information In this example DATA is the Software Branch linked together with LCD Screens and the Keypad Arrow shown above Pressing this Keypad Button X will scroll the LCD to the ROOT Menu P R O D U C T I O N 4 0 R E J 9 7 3 D A T A S T A N D B Y R O M E M B P E R F O R M F E E D 5 3 6 µ S T C P E R M E A T E 8 9 µ S T C R E J 9 7 3 P R O D U C T I O N 4 0 R E J 9 7 3 D A T A S T A N D B ...

Page 12: ...s per millilitre FI Fouling Index GAZ A type of thread a metric thread i e For Example kPa kiloPascals L Litre LCD Liquid Crystal Display LED Light Emitting Diode LPH Litre Per Hour LPM Litre Per Minute LSI Langelier Saturation Index mm millimeter MW Molecular Weight nm nanometer NPT National Pipe Thread NPTF National Pipe Thread Female NPTM National Pipe Thread Male OD Outer Diameter RO Reverse O...

Page 13: ... 0 258 1 L 0 264 US Gal Litres Imperial Gallon 0 215 1 L 0 220 Imp Gal mm inches 0 03937 1 inch 25 4 mm 2 54 cm mm feet 0 0037 1 foot 304 8 mm 30 48 cm kg lb 2 2 1 kg 2 2 lb bar psi 14 7 1 bar 1 b 14 7 psi bar kPa 100 1 bar 100 kPa C F 1 8 C 32 25 C 77 F LPM US Gal per minute 0 264 500 ml min 0 500 LPM 0 132 USG min ...

Page 14: ...ndicate that proper safety equipment has to be used Protective glasses and gloves must be worn This UV RADIATION sticker is used to refer to a position on the water system Cabinet or inside of it where exposure to UV light is possible This DANGER sticker is used to refer to a position on the water system Cabinet or inside of it that could be hazardous This ELECTRICAL GROUND sticker is used to refe...

Page 15: ...his table Elix 20 20 LPH Elix 35 35 LPH Elix 70 70 LPH Elix 100 100 LPH Water temperature 25 C and 10 bar operating pressure up to 12 bar with cold feedwater Water temperature 25 C and 7 bar operating pressure PRODUCT WATER QUALITY Resistivity 5 MΩ cm 25 C typically 10 15 MΩ cm when CO2 is 30 ppm in the Feedwater Conductivity 0 2 µS cm 25 C typically 0 067 0 10 µS cm when CO2 is 30 ppm in the Feed...

Page 16: ...re Elix System In most cases the Membrane and System Recovery values will not be the same number ELIX SYSTEM RECOVERY An Elix System can be operated with a System Recovery between 30 to 70 This value will depend upon several factors including the chemical makeup of the Feedwater System Recovery is defined above See What is RO Recovery The Elix System Recovery is factory pre set If desired the Elix...

Page 17: ...d Valve L RO Permeate Divert Solenoid Valve B Flowmeter M Elix Module C Purge Valve N QGard TL 1 Polishing Pack Optional D Progard TL System Pretreatment Pack O Product Water Resistivity Sensor E Feedwater Pressure Switch P UV Lamp F Feedwater Conductivity Cell Q Loop Resistivity Sensor Optional G RO Pump Motor Pumphead R A10 TOC Monitor Optional H RO Cartridge S Back Pressure Regulator I Flow Con...

Page 18: ... Information Elix 20 35 70 and 100 10 RO CARTRIDGE CONFIGURATION FOR ELIX SYSTEMS The RO Cartridge configuration for each type of Elix Water Purification System is shown below Elix 20 Elix 35 Elix 50 Elix 100 ...

Page 19: ...ws Product Water to exit the Elix System PORT 5 Port 5 is used when a peripheral device is connected to the Elix System This port could be used to allow water to enter the Elix System to the device inside PORT 6 Port 6 is used when a peripheral device is connected to the Elix System This port could be used to allow water to enter the Elix System to the device inside PORT 7 Port 7 is used when a pe...

Page 20: ... to the externally located Inlet Solenoid Valve goes through this grommet The voltage in this electrical cable is either 100 120 or 230 VAC depending upon the voltage supplied to the Elix System GROMMETS 2 THROUGH 7 These grommets are used to allow an electrical cable to go into or out of the Elix System The electrical cable s are used to control or to power peripheral devices The voltage in any o...

Page 21: ...ter Reservoir WHAT ARE THE DIFFERENT TYPES OF PROGARD PRETREATMENT PACKS There are 4 types of Progard TL Packs The choice of a Progard TL is based upon 2 factors These factors are the Total Chlorine Level and the Fouling Index of the Feedwater The type of Progard TL used is determined by the magnitude of these parameters You will be able to choose between a Progard TL1 System Pretreatment Pack Pro...

Page 22: ...ix System The Purge Valve is opened during a Pack Change Mode When it is open air will flow through it The entering air allows any water in the Progard TL to flow into the Elix System and then to drain This allows a wetted Progard TL to be emptied of water This prevents water from flowing out of the Progard TL when it is removed from the Elix System FEEDWATER PRESSURE SWITCH The Feedwater Pressure...

Page 23: ...eate Divert Solenoid Valve is used to divert the RO Permeate Water to drain when the Elix System is in RINSING Mode and during R O RINSE Mode PRESSURE SENSOR The Pressure Sensor is used to measure the water pressure near the Back Pressure Regulator The measured water pressure is displayed on the LCD BACK PRESSURE REGULATOR The Back Pressure Regulator is used to adjust the RO Pump Pressure It also ...

Page 24: ...lectrical power The UV Lamp is used to kill bacteria HOW OFTEN DO I CHANGE THE UV LAMP It is recommended to change the UV Lamp every 5000 hours of operation A timer is incremented every minute the UV Lamp is powered The Elix System displays the number of days before the UV Lamp needs replacement This can be viewed in the DATA portion of the software When the projected lifetime reaches 14 Days the ...

Page 25: ... TL PRODUCT SETPOINT is shown to indicate that the QGard TL Pack should be replaced This message is shown when the resistivity measured after the QGard is less than the Resistivity Setpoint The Resistivity Setpoint can be seen using the LCD This message is not shown after a specific number of days A timer is not used for the QGard TL Pack lifetime LOOP RESISTIVITY SENSOR OPTIONAL Another Resistivi...

Page 26: ...TH THE A10 The A10 uses a small UV light during its TOC analysis mode The A10 UV Lamp needs to be replaced periodically The normal useful lifetime of the A10 UV Lamp is 1 year The Elix System uses a calendar day timer to record the age of the A10 UV Lamp A service message is prompted 14 days before the end of 1 year EXCHANGE A10 UV IN 14 DAYS to remind you to change the A10 UV Lamp The Catalogue N...

Page 27: ...and 100 19 2 7 TECHNICAL SPECIFICATIONS DIMENSIONS SYSTEM HEIGHT WIDTH AND DEPTH The height width and depth of an Elix System are shown here SHIPPING CONTAINER Height 96 cm Width 82 cm Depth 82 cm 44 1 cm 66 2 cm 73 3 cm 5 0 cm 21 4 cm ...

Page 28: ...or any accessories SHIPPING WEIGHT Shipping Weight is defined as a dry system in its shipping container It does not include the Progard TL or any accessories i e an External Booster Pump PROGARD TL WETTED WEIGHT A wetted Progard TL Pack by itself weighs 5 kg OPERATING PRESSURES The Recommended Operating Pressures are shown below An Operating Pressure is defined as the Post Pump Pressure during PRO...

Page 29: ...20 Elix 35 Elix 70 Elix 100 100 VAC 60Hz 500 VA ZLXS80020 500 VA ZLXS80035 500 VA ZLXS80070 500 VA ZLXS80100 100 VAC 50 Hz 500 VA ZLXS70020 500 VA ZLXS70035 500 VA ZLXS70070 500 VA ZLXS70100 120 VAC 600 VA ZLXS60020 600 VA ZLXS60035 600 VA ZLXS60070 600 VA ZLXS60100 230 VAC 500 VA ZLXS50020 500 VA ZLXS50035 500 VA ZLXS50070 500 VA ZLXS50100 MATERIALS OF CONSTRUCTION FOR WETTED COMPONENTS Please co...

Page 30: ......

Page 31: ...K FULL Mode to PRODUCTION Mode This parameter is factory set to 80 When the Elix System is in TANK FULL Mode and water is withdrawn from the Reservoir then the Elix will go to PRODUCTION Mode automatically when the Reservoir is 80 full MANUAL METHOD 1 Start in STANDBY Mode 2 Press the PRODUCTION Keypad Button for about 2 seconds Press PRODUCTION for about 2 3 seconds Lasts for about 16 seconds 3 T...

Page 32: ... to TANK FULL Mode ALARM MESSAGE DISPLAYED 1 Certain Alarm Messages when activated will cause PRODUCTION Mode to stop i e LOW FEED PRESSURE Mode See Section 5 1 for an explanation of the Alarm and Maintenance Messages 2 When the cause of the Alarm Stop Message has been corrected press RESUME to go back to PRODUCTION Mode R I N S I N G 4 0 P R O D U C T 1 3 1 M Ω T C T O C 2 1 p p b D A T A S T A N...

Page 33: ...m 13 1 MΩ cm to 15 0 MΩ cm at 25 C 3 PRODUCTION Mode for an Elix System with a Loop Resistivity Sensor looks like When DIST is shown on the 2nd line then the Resistivity and TOC values are measured from the Loop Water In the example shown immediately above the water in the Loop has a Resistivity of 3 2 MΩ cm at 25 C and a TOC value of 29 ppb In this same example pressing DATA will then show the Re...

Page 34: ...e same for any configuration of an Elix System 4 When the PRODUCT QUALITY is shown press the Bottom Keypad Button to see the RO Cartridge performance 5 When the RO Cartridge performance is shown press the Bottom Keypad Button to see the PRESSURE DATA The RO Cartridge performance is shown here Feedwater pressure is 2 1 bar and the applied RO pressure is 10 0 bar P R O D U C T Q U A L I T Y P R O D ...

Page 35: ...the SETPOINT DATA There are 3 Setpoints in the example shown immediately above The TOC Setpoint is turned off The Elix System Product Water Resistivity will turn on a message if its resistivity is 10 0 MΩ cm The Loop Resistivity Setpoint is 2 MΩ cm A message will be displayed if the Loop Water Resistivity is 2 0 MΩ cm C O N S U M A B L E T I M E R S 2 T O C L A M P R T I M E 1 0 0 D C O N S U M A ...

Page 36: ...n to return to PRODUCTION Mode D A T A P R I N T C U R R E N T V A L U E S 3 0 D H I S T O R Y X S E T P O I N T D A T A T O C O F F P R O D U C T 1 0 M Ω T C D I S T 2 M Ω T C D A T A P R I N T C U R R E N T V A L U E S 3 0 D H I S T O R Y X P R O D U C T I O N 4 5 P R O D U C T 1 5 0 M Ω T C T O C 2 1 p p b D A T A S T A N D B Y ...

Page 37: ...ollow the steps below to see the information listed above Start in STANDBY Mode Press MAINTENANCE Press the DATA Keypad Button The Catalogue Number for the Elix System is shown The Serial Number for the Elix System is shown Press the 4th Keypad Button to access more information about the Elix System The Elix System Installation Date is shown The Software Version is shown The Elix System Production...

Page 38: ...A R E V 1 0 1 0 1 P R O D U C T I O N 1 6 h P R O G A R D S E R N B F 3 N N 8 7 2 5 C I N S T D A T E 0 3 0 9 0 3 R T I M E 5 6 D P R O G A R D S E R N B F 3 N N 8 7 2 5 C I N S T D A T E 0 3 0 9 0 3 R T I M E 5 6 D U V L A M P R T I M E 2 1 5 D T O C L A M P R T I M E 1 0 0 D X P R O G A R D S E R N B F 3 N N 8 7 2 5 C I N S T D A T E 0 3 0 9 0 3 R T I M E 5 6 D Q G A R D S E R N B F 3 N N 5 6 2 ...

Page 39: ...nue until the RO Rejection is within 0 5 of the Memorised Value FLUSH MODE FLUSH Mode occurs when the Inlet Solenoid Valve and the RO Reject Solenoid Valve both are open This allows Tap Water to enter and to sweep away any accumulated impurities on the Feedwater surface of the RO Cartridge FLUSH Mode occurs automatically every 6 hours for 5 minutes in either STANDBY Mode 1st LCD screen shown on th...

Page 40: ......

Page 41: ...ther Pretreatment See Pretreatment Device User Manual for this information See Pretreatment Device User Manual for this information See Pretreatment Device User Manual for this information PrePak L1 Pre System Pretreatment Pack Replacement Change when Progard TL1 or TL2 Pack is changed See information that came with the Pack Progard TL1 or TL2 System Pretreatment Pack Replacement When prompted to ...

Page 42: ... information if necessary Press the MAINTENANCE Keypad Button Press the EXCH PROGARD PACK Keypad Button for a few seconds The Progard TL on the Elix System will automatically be purged of water This helps to prevent water coming out of the ports of the Progard TL when it is removed from the Elix System It will purge for 12 seconds during which time the Pump will operate When the Progard TL is purg...

Page 43: ...gard TL This is done to empty it of air and to hydrate the material inside If you have a Storage Reservoir with a Vent Filter then the LCD will remind you to change the Vent Filter at this time If the Progard TL does not have a chlorine tablet inside of it then the Elix System will go into SYSTEM CLEANING Mode for 5 minutes If the Progard TL does have a chlorine tablet inside of it then the Elix S...

Page 44: ...fore ordering a replacement UV Lamp If the UV LAMP is not replaced on the recommended replacement date then a message UV LAMP EXCH OVERDUE XX DAYS is shown If the UV Lamp is replaced then these messages will not be shown anymore 4 4 HOW TO CHECK AND ADJUST THE PUMP PRESSURE The pump pressure is normally adjusted if the water temperature changes significantly The pump pressure used at different tem...

Page 45: ...automatically be sanitised MANUAL WAY OF SANITISATION CLEANING PORT KIT The RiOs Elix L Cleaning Port Kit is used to introduce the Chemical Sanitant as a dry tablet into the Elix System The Cleaning Port when installed has a removable cap This cap can be unscrewed and a Chemical Sanitant Chlorine Tablet can be placed inside It is not possible to make a Chemical Sanitant in the liquid form and pour...

Page 46: ... form a strong chemical solution when mixed with water This solution can be dangerous for your eyes and your skin Wear Eye Safety Glasses Use Laboratory Gloves and other appropriate safety equipment Press MAINTENANCE Press the Bottom Keypad Button Press Cl2 CLEAN for a few seconds The Elix System will perform a Pressure Test for about 16 seconds The Elix System will go to RO CLEANING Mode It will ...

Page 47: ... R O C L E A N I N G F L U S H I N G R T I M E 5 m i n R O C L E A N I N G R I N S I N G R T I M E 1 m i n T E S T I N G F E E D P R E S S P L E A S E W A I T T E S T I N G F E E D P R E S S P L E A S E W A I T R I N S I N G 4 0 P R O D U C T 1 3 1 M Ω T C T O C 2 1 p p b D A T A S T A N D B Y R I N S I N G 4 0 P R O D U C T 1 3 1 M Ω T C T O C 2 1 p p b D A T A S T A N D B Y P R O D U C T I O N 4...

Page 48: ...AN A OR ROCLEAN B TABLETS Millipore SAS recommends RoClean A or RoClean B as the RO Cartridge cleaner RoClean A is an acidic cleaner The Millipore SAS Catalogue Number for RoClean A is ZWACID012 10 pouches per box RoClean B is a base cleaner The Millipore SAS Catalogue Number for RoClean B is ZWBASE012 10 pouches per box WHICH CLEANER TO USE ROCLEAN A OR ROCLEAN B Contact Millipore SAS for assista...

Page 49: ...e The Cleaning Pouch will form a strong chemical solution when mixed with water This solution can be dangerous for your eyes and your skin Wear Eye Safety Glasses and use Laboratory Gloves and other appropriate safety equipment Press MAINTENANCE Press the Bottom Keypad Button Press pH CLEAN for a few seconds The Elix System will perform a Pressure Test for about 16 seconds The Elix System will go ...

Page 50: ... E 2 5 m i n p H C L E A N I N G S O A K I N G R T I M E 2 6 m i n p H C L E A N I N G F L U S H I N G R T I M E 6 m i n p H C L E A N I N G R I N S I N G R T I M E 5 m i n p H C L E A N I N G F L U S H I N G R T I M E 8 5 m i n p H C L E A N I N G S O A K I N G R T I M E 8 4 m i n p H C L E A N I N G R I N S I N G R T I M E 1 m i n T E S T I N G F E E D P R E S S P L E A S E W A I T T E S T I N G...

Page 51: ...imer is not used for the QGard TL Pack lifetime REMOVING THE USED QGARD TL PACK Place the Elix System into STANDBY Mode See Section 3 1 for more information if necessary Press the MAINTENANCE Keypad Button Press the EXCHANGE POLISHER Keypad Button for a few seconds Unlock the Pack by lifting up the clip that holds the QGard TL Pack in place Gently remove the QGard TL Pack from the Elix System M A ...

Page 52: ...d TL Pack is removed install the new QGard TL Pack with the TAG Sensor at the top The Elix System will automatically identify the QGard TL type by accessing data contained on the Pack TAG Normally the Elix System will recognise and accept the information from the QGard TL Pack Make sure the ports of the new Pack are pushed in After this secure the Pack in place by pushing down on the Pack Locking ...

Page 53: ...ollow these steps below to perform an A10 CLEANING Start in STANDBY Mode Press MAINTENANCE Press the Bottom Keypad Button until TOC CLEAN is shown See the LCD screens below Press TOC CLEAN for 2 3 seconds Press SET A 60 minute timer will be shown S T A N D B Y S E T U P M A I N T E N A N C E P R O D U C T I O N M A I N T E N A N C E D A T A E X C H P R O G A R D P A C K E X C H Q G A R D P A C K M...

Page 54: ...UCTION Mode if the TOC Monitor is measuring the Elix System Product Water The text cl mode will be shown in place of the TOC value This indicates that the TOC CLEANING Mode is taking place T O C C L E A N I N G S E T R T I M E 6 0 M i n R E S E T X M A I N T E N A N C E R O R I N S E T O C C L E A N X M A I N T E N A N C E R O R I N S E T O C C L E A N X M A I N T E N A N C E T 0 D U M P T A N K X...

Page 55: ... P M A I N T E N A N C E P R O D U C T I O N M A I N T E N A N C E D A T A E X C H P R O G A R D P A C K E X C H Q G A R D P A C K M A I N T E N A N C E p H C L E A N C l 2 C L E A N M A I N T E N A N C E D A T A E X C H P R O G A R D P A C K E X C H A N G E P O L I S H E R M A I N T E N A N C E p H C L E A N C l 2 C L E A N M A I N T E N A N C E R O R I N S E T O C C L E A N X M A I N T E N A N C...

Page 56: ...ith the replacement UV Lamp Contact Millipore SAS for a copy of these instructions if you would like to see them before ordering a replacement A10 UV Lamp 4 10 HOW TO CALIBRATE THE A10 TOC MONITOR It is recommended to recalibrate or at least to re verify the A10 TOC Monitoring Device once a year The Elix System will not display any type of message indicating that the A10 needs to be recalibrated T...

Page 57: ...S Accessory is being used with 2 pumps duplexed pumps then no Customised Messages are available When an Elix System is manufactured there are no Customised Messages added during its manufacture HOW MESSAGES ARE DISPLAYED AND OTHER INFORMATION 1 The messages are only displayed during operating modes and in STANDBY Mode For example if you were using the SETUP Menu then you would not see an Alarm Mes...

Page 58: ...act Millipore SAS for assistance CANNOT ACCESS CONSUMABLE INFO The information on the TAG Sensor was not read correctly in PRODUCTION Mode Replace the Progard or accept it Alarm Red LED CHECK A10 COM The communication between the A10 TOC Monitor and the Elix System PC Board is interrupted Contact Millipore SAS Programmable can be set to any type of message Maintenance by Factory Default CHECK ASM ...

Page 59: ...rd TL Pack should be replaced in XX Days The message will go away when a new Progard TL Pack is installed Maintenance Yellow LED EXCH POLISHER PRODUCT SETPOINT The measured Resistivity of the Polisher Product Water is below the Resistivity Setpoint Exchange the Polisher and Product Water Resistivity is Resistivity Setpoint Programmable can be set to any type of message Maintenance by Factory Defau...

Page 60: ...e set to any type of message Maintenance by Factory Default PRODUCT CONDUCTIVITY MINI The measured resistivity from the Polisher or Elix Module is too high 94 9 MΩ cm The displayed resistivity may not show any higher than 18 2 MΩ cm at 25 C Keep the Elix System in PRODUCTION Mode for a few minutes to remove any air that could be trapped in the Resistivity Sensor Contact Millipore SAS for assistanc...

Page 61: ...isher is installed with its TAG Sensor at the top Alarm Stop Red LED and System Stopped QGARD TL POLISHER TYPE ERROR The type of QGard TL Polisher Cartridge has not been identified when it was installed Check the Catalogue Number of the pack using its label Replace if necessary Alarm Red LED QGARD TL POLISHER UNKNOWN The QGard Polisher has not been recognised by the Elix System Verify that the cor...

Page 62: ... Verify that the Level Sensor from the SDS is connected to the Elix System This message will disappear when the water level is 10 full in the SDS Alarm Red LED TOC SETPOINT The measured TOC is the TOC Setpoint Check the value of the TOC Setpoint using the DATA Keypad Button in PRODUCTION Mode Contact Millipore SAS Programmable can be set to any type of message Maintenance by Factory Default TOC ME...

Page 63: ...ld this message is displayed Replace the A10 UV Lamp Maintenance Yellow LED TOO HIGH PUMP PRESSURE The RO Pump pressure is 14 bar for 3 seconds Set the pump pressure lower Contact Millipore SAS for assistance Alarm Stop Red LED and System Stopped UV LAMP EXCH OVERDUE XX DAYS The UV Lamp in the Elix System should have been changed XX days ago The message will go away when a new UV Lamp is installed...

Page 64: ......

Page 65: ...ES Accessory Catalogue Number Comments RiOs Elix L Syst Wall Bracket ZLX0WMBKT Wall Mounting Bracket for an Elix System Wall Bracket PrePak L WBPRPKL01 Wall Mounting Bracket for external PrePak L Water Sensor RiOs Elix L TANKLK001 Water Sensor that connects to an Elix System Water Sensor with cable TANKLK002 Used to add a 2nd Water Sensor RiOs Elix L Cleaning Port Kit ZR00CLNKT Used to add a Clean...

Page 66: ...Ordering Information Elix 20 35 70 and 100 58 6 3 CATALOGUE NUMBERS FOR ALL ELIX SYSTEMS Z X L S 0 5 230 VAC 6 120 VAC 7 100 VAC 50 HZ 8 100 VAC 60 HZ 020 ELIX 20 035 ELIX 35 070 ELIX 70 100 ELIX 100 ...

Page 67: ...0 VAC 10 10 Amp T 5 mm x 20 mm Millipore SAS Spare Part Number FTPF04803 120 VAC 10 10 Amp T 5 mm x 20 mm Millipore SAS Spare Part Number FTPF04803 230 VAC 10 10 Amp T 5 mm x 20 mm Millipore SAS Spare Part Number FTPF04803 1 Use a small flathead screwdriver to open the Fuse Holder This is located immediately above the point where the Power Cord plugs in 2 Remove the entire fuse holder Replace the ...

Page 68: ...allows tubing to be bent 90 Gasket used with the Inlet Solenoid Valve Pipe connection System Identification Plastic Card Other items i e accessories or peripheral devices Fill in the item description ________________________________________________________________________________________________ ________________________________________________________________________________________________ ______...

Page 69: ...cifications for the electrical supply and electrical connections distance to source earth grounded met see the Pre Installation document for these specifications Was the RO Reject Tubing kept to a maximum length of 2 5 meters Was the Progard TL Pack installed such that it is located on front left side of the Elix System it was installed according to the directions given in Section 2 4 in the Insta...

Page 70: ...___________________________________ ___________________________________________________________________________________________________ ___________________________________________________________________________________________________ ___________________________________________________________________________________________________ ________________________________________________________________...

Page 71: ...ssary FEEDWATER PIPING CONNECTION REQUIRED The Feedwater should be supplied in a pipe that terminates in a 1 2 inch Male GAZ connection This is the type of connection needed for the Inlet Solenoid Valve A 1 2 inch NPTM or 1 2 inch BSP connection can also be used FEEDWATER CHEMISTRY REQUIRED Potable Mains Tap Water Conductivity 2000 µS cm 25 C Calcium 300 ppm as CaCO3 when System Recovery is 50 pH ...