Message

General Cause

Corrective Action

Can not find file or 
credentials are incorrect

Does not connect to  
Wi-Fi

® 

network

Check or insert the “credential file” on USB thumb drive.  
See 

Creating a Wi-Fi

®

 Configuration File, p .9 .

Network scan timeout

No 2.4 GHz network(s)  
within range

Go within range of a 2.4 GHz wireless network  
and try again.

Cannot find file 
(networkname.config)

No USB thumb drive inserted
or USB thumb drive is not  
properly formatted, 
or proper config file  
not present on drive

Ensure proper config file is located on FAT32 formatted  
USB thumb drive and try again. See

 Wi-Fi

®

 Network, p .8 

to create a config file.

Invalid Config 
(networkname.config  
line #)

Config file not correct

See 

Creating a Wi-Fi

®

 Configuration File, p .9 .

Network scan timeout

No 2.4 GHz network(s)  
within range

Go within range of a 2.4 GHz wireless network and  
try again.

WIFI Config (networkname.
config) missing paraphrase

Config file not correct

Recreate config file. See 

Creating a Wi-Fi

®

 Configuration 

File, p .9 .

Wrong or missing update  
file on USB thumb drive

No USB thumb drive 
inserted, drive is not properly 
formatted, 
or corrupt update file on drive

Ensure proper update file is located on FAT32 formatted USB 
thumb drive and try again. See 

USB Thumb Drive Formatting, 

p .8 .

Software update failed

Corrupt update file on drive

Ensure proper update file is located on FAT32 formatted USB 
thumb drive and try again. See 

USB Thumb Drive Formatting, 

p .8 .

Software update blocked  
on battery low

Insufficient battery to  
perform update

Recharge Scepter™ 3.0 and try again.

File(s) did not print

Printer malfunction 
or not connecting to printer
or Low Battery

Confirm printer is on, and there is no paper jam.  
See 

Bluetooth

®

 Enabled Printer, p .7 

to establish  

new printer connection.

Confirm WiFi

®

 signal is on, WiFi

®

 icon should be visible  

on screen. See 

Wi-Fi

®

 Network, p .8 

to establish a new  

network connection.
Recharge Scepter™ 3.0 and try again.

Connection was 
unsuccessful

Printer not ready

Printer spooling 
unsuccessful

Printer Scan timeout

Printing unsuccessful...
Make sure printer is on  
and within range

Self test failed results

Scepter™ 3.0 malfunction

Replace sensor and retry.  
If unsuccessful, contact Tech Service.

 

Notice

We provide information and advice to our customers on application technologies and regulatory matters to 

the best of our knowledge and ability, but without obligation or liability. Existing laws and regulations are 

to be observed in all cases by our customers. This also applies in respect to any rights of third parties. Our 

information and advice do not relieve our customers of their own responsibility for checking the suitability 

of our products for the envisaged purpose.
The information in this document is subject to change without notice and should not be construed as a 

commitment by the manufacturing or selling entity, or an affiliate. We assume no responsibility for any 

errors that may appear in this document.

Summary of Contents for PHCC30000

Page 1: ...perates as MilliporeSigma in the U S and Canada The life science business of Merck KGaA Darmstadt Germany operates as MilliporeSigma in the US and Canada SigmaAldrich com User Guide Scepter 3 0 Handhe...

Page 2: ...r computer easily No more cables The new Charging Station stores Scepter 3 0 out of the way when not in use freeing up valuable hood or lab bench space This system is intended for research use only Co...

Page 3: ...nstrument to vibrations Vibrations may cause instrument malfunction or damage Do not autoclave or expose to high temperature To avoid damage to internal electronic components do not spray instrument w...

Page 4: ...s the sensor Tech Service Port For Service use only USB Port Connects instrument to USB thumb drive Ergonomic Hook Allows comfortable handling of the instrument Secures the instrument to charger Contr...

Page 5: ...the Recharging screen when placed in the Charging Station the following steps must be followed to start charging the Scepter 3 0 battery 1 Place Scepter 3 0 into the Charging Station 2 While on the s...

Page 6: ...kable surface 1 Use double sided mounting tape 2 Remove tape liner from one side of tape and affix to back of mounting plate as shown below 3 Remove remaining tape liner and press firmly onto glass me...

Page 7: ...14 16 Edit Time Hours Minutes Seconds 14 16 34 Press OK to accept 01 Feb 18 14 16 Edit Time Hours Minutes Seconds 14 16 34 Press OK to accept 1 Scroll to Edit Time and press the OK button 2 Scroll to...

Page 8: ...nd select Properties If the File system is identified as FAT32 no formatting is needed Continue to Creating a Wi Fi Configuration File NOTE This view will also show how much space is available on your...

Page 9: ...file from the computer onto the USB thumb drive Once the file has loaded remove the USB thumb drive from the computer port Connect to Wi Fi Network 1 From the Main Menu select Settings and press the...

Page 10: ...were last set by the user either during a count or during post count data management 01 Feb 18 14 16 Menu Home Histograms Settings Help 01 Feb 18 14 16 Settings Back Date Time Bluetooth Wifi Gating M...

Page 11: ...epter 3 0 instrument Do not use sensor if visibly damaged Attach the Sensor 1 When the Scepter 3 0 is powered on it will show Attach the Sensor to begin 2 Holding the sensor with the sensor size numbe...

Page 12: ...net is recommend use low vacuum 5 Hg Cell Bead and Particle Suspensions Materials Required Cell sample Scepter 3 0 Test Beads Cat No PHCC3BEADS or particle suspensions Note Some particle suspensions...

Page 13: ...ion and diameter or volume will be displayed on the screen Press the OK button to save up to 999 histograms in the Scepter 3 0 memory and open the Histogram Explorer Press the OK button to return to H...

Page 14: ...03 Dec 17 Count 0 900 36 24 30 18 12 6 Dia in m 3 342e5 15 685 mL m Conc Dia 01 Feb 18 14 16 Histogram Edit Gates High gate Done 26 575 Low gate 12 585 Count 0 900 36 24 30 18 12 6 Dia in m Switch to...

Page 15: ...wirelessly Materials Required Wireless Network broadcasting at 2 4 Ghz PC Mac or UNIX Computer connected to the same Wi Fi Network as Scepter 3 0 Scepter 3 0 Instrument Follow Wi Fi Network p 8 direct...

Page 16: ...MS Press the OK button from any histogram on screen to open the Histogram Explorer menu 01 Feb 18 14 16 Menu Home Histograms Settings Help 01 Feb 18 14 16 Histogram Test 3 03 Dec 17 Count 0 900 36 24...

Page 17: ...o printer Help Screens Tutorials 1 From the Menu screen scroll to Help press the OK button 2 Scroll to the topic of interest press the OK button 01 Feb 18 14 16 Scepter Software 4 6 01 Feb 18 14 16 Me...

Page 18: ...ration of cell sample is within recommended operating range See Cell Bead and Particle Suspensions Test sample using Scepter 3 0 Test Beads Cat No PHCC3BEADS to ensure unit operating properly for conc...

Page 19: ...y needs to be re charged Place Scepter 3 0 on charging stand to recharge the instrument See Battery p 5 Battery is not charging Ensure that the unit is correctly positioned in charging stand An animat...

Page 20: ...ressing of OK button Fully submerge sensor in sample before pressing OK button Air bubble in sample Mix sample gently to avoid air bubbles Sensor lifted from sample before screen displays Sample Loade...

Page 21: ...ed on FAT32 formatted USB thumb drive and try again See USB Thumb Drive Formatting p 8 Software update blocked on battery low Insufficient battery to perform update Recharge Scepter 3 0 and try again...

Page 22: ...ntal Conditions Temperature 15 C to 25 C Maximum Relative Humidity 20 80 non condensing Storage Conditions Temperature 15 C to 30 C Maximum Relative Humidity 20 80 non condensing Scepter 3 0 Sensor Ma...

Page 23: ...p E GP Sterile Vacuum Filtration System Millipore Express PLUS PES 0 22 500 38 12 SEGPU0538 0 22 500 45 12 SEGPU0545 0 22 1000 38 12 SEGPU1138 0 22 1000 45 12 SEGPU1145 Steritop E GP Sterile Funnel le...

Page 24: ...Merck KGaA Darmstadt Germany and or its affiliates All Rights Reserved Global Registration Marks of Conformity USA TFB 1003 EUROPE C TAIWAN JAPAN R209 J00212 SOUTH KOREA R C Emd PHCC30000 RUSSIA BELAR...

Reviews: