HistoDream EW
MM164-001-Operator Manual
40
5. CLEANING UP
The procedure described in this chapter must be done at least once a day or following the
regulations in use in the laboratory.
For reasons of safety, any operation without the required personal protection equipment
is to be avoided.
5.1. Safety precautions
The personnel involved in cleaning up HistoDream EW and its optional parts (if any) must wear personal
protective equipment, specifically:
•
HI-RISK type single-use protection gloves (100% latex or equivalent material). Personal protection
system compliant with standard EN 374 and Directive 89/686/EEC.
•
Antibacterial mask for surgical/operating theatre use and white coat. Personal protection system
compliant with standard EN 149 Grade FFP3D and Directive 89/686/EEC.
5.2. HistoDream EW cleaning up
Strictly follow the instructions below to properly perform the cleaning up procedure.
For further information, please email our specialists at:
Do NOT remove the side panels of the unit during the cleaning up operations.
For cleaning up, do not use procedures other than that described in this chapter.
Before every operation disconnect the power supply line. Do not use direct or high-
pressure water to clean the unit.