HistoDream EW
MM164-001-Operator Manual
11
1.3.3. HistoDream TM, Thermal Module
MODEL
REF 500003-TB
REF 500003
•
Electrical
connections
To connect only with: HistoDream DM,
Dispensing Module REF 500002-TB
100-120V
50-60Hz
To connect only with: HistoDream DM,
Dispensing Module REF 500002
220-240V
50-60Hz
For both HistoDream TM, Thermal Module models:
•
Dimensions
Width
Depth
Height
340mm (13.4’’)
623mm (24.5’’)
395mm (15.6’’)
•
Weight
16kg (35.3lb)
•
Thermal chamber for moulds
Width
Depth
Height
- Dimensions
269mm (10.6’’)
510mm (20.1’’)
88mm (3.5’’)
- Capacity
2 removable trays with capacity for 2 racks (200 cassettes)
Without removable trays capacity for 3 racks (300 cassettes)
•
Thermal chambers for
cassettes
Width
Depth
Height
- Dimensions
275mm (10.9’’)
170mm (6.7’’)
180mm (7.1’’)
- Capacity
400 moulds
1.4. Transport and storage conditions
- Transport and storage conditions:
from -20°C (-4°F) to +50°C (122°F)
- Operating conditions:
from 15°C (59°F) to 26°C (78.8°F) (at a 60% maximum relative humidity)
Altitude up to 2000m above sea level
Indoor use only
Before switching on the unit (when coming from a storage chamber), allow the instrument
to reach operating environment conditions (at least half an hour).
Repair or maintenance work is usually performed at the site of installation. If this is not possible for some
particular reasons, the instrument may be returned to Milestone. The contact address can be found at the
end of this instruction manual.