Message displayed
Cause
Detection/remedy
E
Input voltage
incompatible
Voltage of power supply is
too high.
Reader Unit switches off automatically after
20 seconds.
Use original power supply
F
Wrong SD-card inserted!
SD-card!
:
Measurement w/o storage of
data possible
- or -
Insert SD-card with correct ID
XXXXXXXXXX!
Measurement starts wit-
hout SD-card. SD-card not
designated for use with the
implant is employed during
measurement
Insert the SD-card designated for use with
the shunt sensor into the Reader Unit. Use the
identification number (ID) to assign the cor-
rect SD-card to the implant (see patient pass).
Wrong SD-card inserted!
:
Measurement w/o storage of
data possible
-or -
Insert SD-card with
correct ID XXXXXXXXX!
SD-card with ID which dif-
fers to that of the implant
Insert the SD-card designated for use with
the shunt sensor into the Reader Unit. Use the
identification number (ID) to assign the cor-
rect SD-card to the implant (see patient pass).
Wrong implant
-restart measurement!
:
Switch to another implant du-
ring continuous measurement
not possible!
Once continuous measure-
ment has started, data are
received from a different
implant
Increase the distance between the two im-
plants
G
Unit-temperature incorrect
:
Unit temperature from 10°C to
40°C valid!
Temperature in Reader Unit
exceeds calibrated interval
Reader Unit can only be used at a device tem-
perature of 10°C to 40°C.
Current measurement is
interrupted
I
Internal voltage
incompatible
Internal voltage of device
too high/too low
Reader Unit switches off
automatically after 20 seconds.
Contact Technical Service
S
SD-card faulty!
:
Measurement w/o
storage of data possible!
SD-card not writable or
readable (dirt, corrosion,
faulty contacts)
Use of implant‘s internal calibration data.
Data are not saved
SD-card faulty!
SD-card not readable (dirt,
corrosion, faulty contacts)
Check SD-card for damage or dirt
SD-card inserted!
Restart measurement!
:
Storage of reading after restart
of measurement possible!
Measurement starts without
SD-card. SD-card belonging
to the implant is inserted
during measurement
Start measurement again
Message displayed
Cause
Detection/remedy
SD-card inserted!
Restart measurement!
:
Storage of reading after restart
of measurement possible!
Measurement starts without
SD-card. SD-card belonging
to the implant is inserted
during measurement
Start measurement again
SD-card missing!
:
Insert SD-card!
No SD-card inserted in the
data manage-ment mode
Insert SD-card
SD-card missing!
:
Measurement w/o storage of
data possible
- or -
Insert SD-card with
correctID XXXXXXXXXX!
SD-card not inserted
Insert the SD-card designated for use with
the shunt sensor into the Reader Unit.
Use the identification number (ID) to assign
the correct SD-card to the implant
(see patient pass).
SD-card not readable!
:
Measurement w/o
storage of data possible!
Insertion of
wrongly formatted or unfor-
matted
SD-card
Insert the SD-card designated for use with
the shunt sensor into the Reader Unit.
Use the identification number (ID) to assign
the correct SD-card to the implant
(see patient pass)
SD-card not readable!
Measurement with
storage of data
possible!
No SD-card
- or -
SD-card ID not readable
- or -
SD-card contains no cali-
bration data
Insert the SD-card designated for use with
the shunt sensor into the Reader Unit.
Measurement with data storage possible, but
with limited accuracy.
Measurement with data storage possible, but
with limited accuracy
SD-card memory full Option to
measure w/o saving!
Memory map on SD-card
exhausted (100%) during
continuous measurement
Delete any data that are no longer needed,
measurement possible without storage
SD-card memory full
Memory map on SD-card
exhausted (100%) during
continuous measurement
Delete any data that are no longer needed
SD-card memory almost full
Memory map on SD-card
almost exhausted (99%)
during continuous measure-
ment or fast measurement
Delete any data that are no longer needed
System error
Ambient pressure not readable
Barometric (BARD) pressure
not readable
Device function locked. Deactivate and reac-
tivate the device.
If the error persists, please contact Technical
Service