INSTRUCTIONS FOR USE
INSTRUCTIONS FOR USE
|
GB
GB
Phase I
· Remove any visible residues with a disposa-
ble disinfectant wipe.
· Wipe all surfaces of the optically
clean product with a fresh, disposable disin-
fectant wipe.
· Observe the specified contact time (1 minute
minimum).
Inspection
· After each complete cleaning/disinfection
cycle, check the product for damage.
· Set aside the product if it is damaged.
Storage
· Store reprocessed products protected from
dust in a dry, dark and cool room which is
exposed to as little bacterial contamination.
MAINTENANCE/CALIBRATION
The
Reader Unit
requires no maintenance.
CALIBRATION OF BARD SENSOR
The Reader Unit is equipped with a barometric
pressure sensor (BARD sensor). To ensure that
pre-defined tolerance limits are upheld, calibra-
tion is necessary on an annual basis.
For the purposes of calibration, the device
should be sent each year to the following
address:
Service addresses:
Christoph Miethke GmbH & Co. KG
Ulanenweg 2
14469 Potsdam / Germany
Phone:
+49 (0) 331 62083-0
Fax:
+49 (0) 331 62083-40
E-Mail: [email protected]
Note: When undergoing calibration, the device
will also be subjected to thorough functional
and technical safety inspection.
TROUBLESHOOTING
If an error occurs, it will be displayed on the
screen.
Example of an error message:
error message
Fig. 52
Additional information can be retrieved by pres-
sing
Return to the previous menu by
pressing
.
MALFUNCTIONS DISPLAYED WITH AN
ERROR MESSAGE
Note: The table is sorted alphabetically by
the initial letter of the error message
appearing on the display
Message displayed
Cause
Detection/remedy
A
Battery flat
Battery capacity
exhausted (0%)
All data are saved after 2 min. The Reader
Unit switches off automatically.
Connect power supply.
Battery voltage incompatible -
use original power supply
Battery voltage of Reader
Unit is too low
Reader Unit switches off automatically after
20 seconds.Connect original power supply.
Low battery voltage
Battery voltage too low
The backlight switches off after 3 seconds.
Connect power supply.
Current measurements will not be inter-
rupted.
B
Flashing, crossed-through an-
tenna symbol
:
Antenna error
Antenna faulty
Deactivate and reactivate the device.
If the error persists, please contact Technical
Service.
Flashing, crossed-through an-
tenna symbol
:
Antenna not connected
Antenna not connected at
start of measurement.
-or-
Antenna was disconnected
during measurement.
Connect antenna.
Measurement starts again.
-or-
Connect antenna. Measurement continues.
Flashing, crossed-through an-
tenna symbol
:
No communication
Data recording interrupted
during continuous measure-
ment (telemetric link dis-
rupted)
Measurement resumes automatically once
communi-cation is restored
Flashing, crossed-through an-
tenna symbol
SD-card has been removed,
measurement possible
SD-card was removed duri-
ng continuous
measurement.
Insert the SD-card.
Start measurement again
D
Data error!
:
File opening not possible
Validation of the file failed
File cannot be opened
Continues key activation
key pad error
Permanent
actuation of a key >60 se-
conds
Release the key
Pressure values out of range
Pressure data measured
from implant not plausible
- physiologically meanin-
gless data
Measurement continues.
If the error persists, please contact Technical
Service