C. AWK – Automatic mobile unit
recognition
^
Select operating level C.
^
Push the coded mobile unit into
place.
,
Make sure that there are no
small metallic objects or instrument
parts attracted to the magnetic strip,
in particular to the underside of it.
Any metallic objects on the strip can
result in the coding being incorrectly
read.
^
Close the door.
,
Before starting a programme it is
absolutely essential
to check that
the programme shown in the display
is the correct on for the mobile unit
being used.
Otherwise inadequate cleaning or
disinfection could be the result.
Please make sure, that the places
assigned for the programmes using
Automatic mobile unit recognition
are not changed around arbitrarily.
^
Press the Start button
6
.
When processing medical devices,
any changes made to programmes
or dispensing systems must be
documented in a log book kept with
the machine (in accordance with the
Medical Devices directive
93/42/EWG). The machine's
cleaning and disinfecting standards
must also be re-validated.
Programme sequence
The programme will start automatically
as soon as the Start button has been
pressed. It is finished when
PROGRAMME END appears in the
display and the background lighting
flashes (press any button to switch off
the flashing lights).
See "System functions" in the
programming manual for further
information on switching off the flashing
signal.
Detailed information on programme
sequences is given in the appendix
of the Programming Manual.
The display background lighting
goes out automatically after approx.
15 mins. To bring it back on again,
press one of the buttons.
,
On machines with a built-in
printer you must not change the
colour ribbon cartridge or the paper
roll whilst a programme is in
progress.
Switching off
^
Press and release the
I-0
button.
^
Turn off the stopcocks.
Operation
37