The washer-disinfector can be fitted
with various mobile units and baskets,
each adapted to the type and shape of
the items that need to be cleaned and
disinfected; they in turn can be
equipped with a wide variety of
modules and inserts.
Mobile units, baskets and inserts
should be properly selected to
match the application involved.
On the following pages you will find
loading examples and additional
information.
Before EVERY program start,
conduct the following visual
inspection:
– Are the items properly sorted, loaded
and connected in the machine?
– Are the spray arms clean and can
they rotate freely?
– Is the filter combination free of
coarse debris (remove any coarse
material and clean the coarse, fine,
flat and micro-fine filters as needed)?
– Is the adapter properly connected to
the water supply for the spray arms
and nozzles?
– Are the process chemical dispensers
full (detergent, neutralizer, etc.)?
After EVERY program completion,
conduct the following visual
inspection:
– Inspect the labware for cleanliness.
– Have any hollow items been
dislodged from their proper position
on the nozzles?
,
Any items that were dislodged
from the adapters during a program
cycle must be washed again.
– Are the cavities and channels in
hollow items cleaned through?
– Are the nozzles and connections
firmly attached to the mobile
units/inserts?
Protein testing
After washing, items should be
checked at random for proteins.