Loading instructions
Important:
The "Warnings and Safety instruc-
tions" pages must be observed.
WARNING - PERSONAL INJURY/
EQUIPMENT DAMAGE HAZARD:
Internal cleaning and disinfecting to
a measurable standard can only be
achieved with instruments that can
be dismantled and where the jets,
sleeves and adapters are used ap-
propriately. A special examination,
over and above a visual one, of the
cleanliness of the inner area of the
instruments should be made. MIS in-
struments should preferably be pro-
cessed using the DISIN VARIO
93°C-10 MIN or DISIN 93°C-10 MIN
program with recommended deter-
gents.
^
Put 5 mm and 10 mm Ø shafts into
the appropriate sleeves (E-442 or
E-443).
^
Do not load the lower levels with
more inserts than as shown above in
the example (from left to right
E-457, E-460, E-451)
^
Trocars of a wider diameter will be
cleaned more effectively in the front
or the rear position of the row of jets.
Other accessories and fitting not pic-
tured:
^
E 442
5 mm Ø injection sleeve
^
E 443
10 mm Ø injection sleeve
^
E 445
caps for E 442 sleeves
^
E 446
caps for E 443 sleeves
^
E 447
Adapter for Luer lock
(female)
^
E 452
Injector jet Ø 2.5 x 60 mm
^
E 362
Blanking screw
^
E 444
Insert for fiber optic
(cold-light) cable or
suction hose
MIS injector mobile unit E 450
20