Loading the washer/disinfector
WARNING - FALL HAZARD: To pre-
vent falls, keep floors dry by immedi-
ately wiping up any spilled liquids in
washer/disinfector loading and un-
loading areas.
CAUTION: Load baskets and inserts
into washer/disinfector in accor-
dance with the instructions provided.
Care should be taken to remove (den-
tal) cement or amalgam from the instru-
ments. However, where necessary, the
instruments can be cleaned manually
after thermal disinfection (following
health and safety regulations) with mini-
mized danger.
Make sure the spring loaded coupling
engages correctly when the upper bas-
ket is inserted into the washer/
disinfector. There should be a tight fit
between the coupling and the water in-
let connector at the top of the chamber.
If there is not, adjust the adapter as fol-
lows:
^
Loosen the holding ring on the upper
basket.
^
Push up the adapter so it is 1/4’’
higher than the water inlet connector,
and tighten the holding ring.
Adjustable upper basket
Adjustable upper baskets can be used
in this machine. They can be raised or
lowered by
3
/
4
" (2 cm) from the initial
setting. In the middle position, the up-
per basket will accommodate:
^
instruments, utensils, accessories
and inserts up to 7
1
/
2
" (19 cm) in
height
while the lower basket can accommo-
date:
^
instruments, utensils, accessories
and inserts up to 9
13
/
16
" (25 cm) in
height.
For further details regarding con-
tainer height combinations, see table
below.
Top
basket
position
Top basket height
Bottom basket
height
(cm)
(inches)
(cm)
(inches)
top
17
6
11
/
16
27
10
5
/
8
middle
19
7
1
/
2
25
9
13
/
16
bottom
21
8
1
/
4
23
9
1
/
16
To adjust the upper basket
^
Pull out the upper basket until a resis-
tance is felt, lift from the runners and
remove.
Unscrew the knurled nuts on both sides
of the basket and reposition as neces-
sary.
Loading the washer/disinfector