26-0002-0005/0 (en) (2022)
11
2.2 How to Use the Stroke Application & Programming Software (SAPS)
The SAPS has been pre-installed on the commercially available laptop provided.
1. Position the laptop within 1 m of the IPG implanted in the patient’s left chest.
2. Turn the laptop on (Typically the ON button is directly below the laptop screen on the front
left, as shown above in Figure 2.1:, although it may also be on the front right or side of the
computer; press down).
3. The laptop will power up in the “
microtransponder
” login state. Log into the computer
using the password provided.
CAUTION: If the laptop does not automatically launch the SAPS, then please reboot your
computer and re-launch SAPS (see Section 13 Troubleshooting ). If this does not
correct the issue or you have additional problems, please contact MicroTransponder per
Section 18: Information and Support.
.
4. The SAPS loaded onto the laptop will then launch to the Startup Menu Screen (
Welcome
menu) (see Figure 2.22). Before proceeding to the next section, ensure that you have
arrived at this Startup Menu Screen.
Figure 2.22: Startup Menu Screen—Welcome Menu
CAUTION: If the computer does not display the Startup Menu Screen, see Section 12:
Precautions, Warnings, Contraindications, and Hazards and Section 13:
Troubleshooting before proceeding.
For proper operation and communication, the USB plug of the WT must be plugged into a USB
port of the laptop. The WT should be positioned within 1 m of the implanted IPG for proper
communication. The
Interface
indicator on the SAPS user interface screen will be green in color
Summary of Contents for 2100
Page 22: ...26 0002 0005 0 en 2022 22 Figure 4 13 Frequency Figure 4 14 Train Duration ...
Page 35: ...26 0002 0005 0 en 2022 35 Figure 5 10 Magnet Swipe ...
Page 44: ...26 0002 0005 0 en 2022 44 Figure 8 2 Patient Log File Export Log ...
Page 45: ...26 0002 0005 0 en 2022 45 9 INTENTIONALLY BLANK ...