![MicroTransponder 2100 Manual Download Page 50](http://html.mh-extra.com/html/microtransponder/2100/2100_manual_1789717050.webp)
26-0002-0005/0 (en) (2022)
50
13 T
ROUBLESHOOTING
This section provides troubleshooting instructions in 3 parts:
Section 13.1: General
Section 13.2: Troubleshooting at Patient Follow-up Visits
13.1 General Recommendations
13.1.1 Non-responsive Laptop
The SAPS laptop will power up in the
Welcome
menu of the SAPS application. If for any reason
this does not occur, power off the laptop and power it back on. If the problem persists, contact
MicroTransponder per
Section 18: Information and Support.
13.1.2 Other Hardware Issues
For all other hardware issues for the laptop, contact MicroTransponder per
Section 18:
13.2 Troubleshooting at Patient Follow-up Visits
13.2.1 Telemetry Lost in Therapy Visits
An example of an error message following a communication problem is shown in Figure 13.1.
Figure 13.1: Telemetry Lost
A communication problem can cause an error message (such as the “Telemetry Lost” message
shown in Figure 13.1) during interrogation, checking impedance, programming stimulation
parameters, or while performing any other therapy required communications.
The most common causes of communication problems are connection issues. Performing these
steps may restore communication:
1. Verify that the WT is properly connected to the laptop. The USB plug from the WT should
be inserted into one of the laptop’s USB ports. Figure 2. shows the indicator for proper
connection.
2. Verify that the patient and IPG are within 1 m of the WT.
3. Verify that the programming problem is not a result of EMI from nearby electrical or
magnetic equipment. Examples of possible sources of EMI are laptop displays, portable
telephones, fluorescent lighting, OR lights, and magnetic pads for surgical instruments.
4. Verify that the programming problem is not a result of the IPG or WT being placed on a
metal table.
Failure of the IPG and WT to clearly communicate with each other at any time can be attributed
to several factors, such as:
• WT not properly connected to laptop
Summary of Contents for 2100
Page 22: ...26 0002 0005 0 en 2022 22 Figure 4 13 Frequency Figure 4 14 Train Duration ...
Page 35: ...26 0002 0005 0 en 2022 35 Figure 5 10 Magnet Swipe ...
Page 44: ...26 0002 0005 0 en 2022 44 Figure 8 2 Patient Log File Export Log ...
Page 45: ...26 0002 0005 0 en 2022 45 9 INTENTIONALLY BLANK ...