![MicroTransponder 2100 Manual Download Page 54](http://html.mh-extra.com/html/microtransponder/2100/2100_manual_1789717054.webp)
26-0002-0005/0 (en) (2022)
54
Outer Material
PMMA plastic (plexiglas - black)
Size
~8.5 cm × 5.5 cm × 1.5 cm; no sharp corners
Cable Length
~2 m
Weight
~108 g
Connector Type
USB
Communication Distance
≤1 m
Communication Band
Medical Implantable Band (MICS, ~403 MHz)
Table 16.1: Characteristics of the Wireless Transmitter
16.4 Stroke Application & Programming Software (SAPS)
The SAPS software was developed by MicroTransponder to manage IPG therapy and therapy
information. The SAPS has the following capabilities:
• Allows the physician to set the IPG stimulation parameters
• Allows the physician to check the status of the IPG battery level
• Allows the physician to check the Lead impedance
• Allows the physician to record the rehabilitation task information
• Stores the therapy session history for review at a later time
• Requires the physician to use a SAPS password to access implant programming and
therapy functions
• Allows users without the SAPS password (but with the laptop password) to access patient
log and therapy information
The software is delivered to the physician pre-installed on a commercially available laptop.
See
Section 3: Using the Software (SAPS)
for more information on use of the SAPS.
16.5 Other Applicable MicroTransponder
®
Products
• Model 1001 IPG
• Model 3000 Lead
16.6 Applicable Commercial Devices for Use with MicroTransponder IPG
A commercial magnet with a pull force rated at 4.5 kg or greater is sufficient to close the reed
switch within the IPG and stop stimulation.
Summary of Contents for 2100
Page 22: ...26 0002 0005 0 en 2022 22 Figure 4 13 Frequency Figure 4 14 Train Duration ...
Page 35: ...26 0002 0005 0 en 2022 35 Figure 5 10 Magnet Swipe ...
Page 44: ...26 0002 0005 0 en 2022 44 Figure 8 2 Patient Log File Export Log ...
Page 45: ...26 0002 0005 0 en 2022 45 9 INTENTIONALLY BLANK ...