26-0002-0005/0 (en) (2022)
34
5.2 Magnet Mode
CAUTION: Prior to all patient MRI procedures, program the Magnet Mode to Off.
Magnet Mode
set to
On
enables you to initiate a therapy session with the use of a magnet. To
begin this therapy session:
1. In the
Program Implant
menu, ensure that the button adjacent to
Magnet Mode
is
On
.
2. Select the
Magnet Mode Duration
.
3. Program the settings into the IPG by selecting the
Download Settings to IPG
button.
Once the appropriate device settings for
Magnet Mode
are downloaded to the IPG,
stimulation may be triggered using a magnet swipe without further communication from the
SAPS.
NOTE: During a patient’s first therapy session, give the patient the magnet and review this
section and the Microtransponder
®
Vivistim
®
Paired VNS
™
System Quick Start Guide for
Patients with them.
4. Remove the magnet cover, if applicable (see Figure 5.9 below).
Figure 5.9: Magnet with Cover Attached (A) and Detached (B)
5. To begin stimulation, swipe the magnet across the patient’s chest where the IPG is located
(see Figure 5.). This will cause the VNS session to begin for the programmed
Magnet
Mode Duration
.
NOTE: The magnet need only be swiped across the patient’s chest where the IPG is located to
trigger stimulation. Holding the magnet over the IPG for greater than 30 seconds stops the
stimulation until the magnet is removed.
A
B
Magnet
Magnet Cover
Summary of Contents for 2100
Page 22: ...26 0002 0005 0 en 2022 22 Figure 4 13 Frequency Figure 4 14 Train Duration ...
Page 35: ...26 0002 0005 0 en 2022 35 Figure 5 10 Magnet Swipe ...
Page 44: ...26 0002 0005 0 en 2022 44 Figure 8 2 Patient Log File Export Log ...
Page 45: ...26 0002 0005 0 en 2022 45 9 INTENTIONALLY BLANK ...